Perclose ProGlide embolization as a complication: case report and review of literature.
Future Cardiol
; 17(7): 1193-1197, 2021 10.
Article
em En
| MEDLINE
| ID: mdl-33448229
ABSTRACT
Background:
Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. Case We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose ProGlide Suture-Mediated Closure was used for closure. The closure was complicated by a separation of the ProGlide device resulting in the migration of the footplate to the descending aorta. Correction required snare retrieval via radial access, and the patient recovered without complications.Discussion:
We highlight an important complication of the Perclose ProGlide Suture-Mediated Closure device that is rare but important to know when performing endovascular closures with this device.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Procedimentos Endovasculares
/
Dispositivos de Oclusão Vascular
Limite:
Aged80
/
Humans
/
Male
Idioma:
En
Revista:
Future Cardiol
Ano de publicação:
2021
Tipo de documento: