The advantages and disadvantages of the novel fourth-generation cryoballoon as compared to the second-generation cryoballoon in the current short freeze strategy.
Miyazaki, Shinsuke; Hasegawa, Kanae; Mukai, Moe; Aoyama, Daisetsu; Nodera, Minoru; Shiomi, Yuichiro; Tama, Naoto; Ikeda, Hiroyuki; Ishida, Kentaro; Uzui, Hiroyasu; Tada, Hiroshi.
Afiliação
  • Miyazaki S; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan. mshinsuke@k3.dion.ne.jp.
  • Hasegawa K; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Mukai M; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Aoyama D; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Nodera M; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Shiomi Y; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Tama N; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Ikeda H; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Ishida K; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Uzui H; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
  • Tada H; Department of Cardiovascular Medicine, Faculty of Medical Sciences, University of Fukui, 23-3 Shimo-aiduki, Matsuoka, Eiheiji-cho, Yoshida-gun, Fukui, 910-1193, Japan.
J Interv Card Electrophysiol ; 63(1): 143-152, 2022 Jan.
Article em En | MEDLINE | ID: mdl-33575920
ABSTRACT

BACKGROUND:

The novel fourth-generation cryoballoon (4th-CB) is characterized by a shorter-tip that potentially facilitates better time-to-isolation (TTI) monitoring. We sought to clarify the advantages and disadvantages of the 4th-CB compared to the second-generation cryoballoon (2nd-CB) in pulmonary vein isolation (PVI).

METHODS:

Forty-one and 49 consecutive atrial fibrillation patients underwent 2nd-CB and 4th-CB PVIs using 28-mm balloons and short freeze strategies. When effective freezing was not obtained, the CB was switched to the other CB.

RESULTS:

The rate of successful PVIs was significantly higher for 2nd-CBs than 4th-CBs (162/162[100%] vs. 178/193[92.2%] PVs, p < 0.0001). The difference was significant for lower PVs, especially right inferior PVs (RIPVs)(p = 0.005). In a total of 15 PVs in 11 patients, 4th-CBs were switched to 2nd-CBs, and 14/15(93.3%) PVs were successfully isolated. The balloon temperature tended to reach -55℃ more frequently with 2nd-CBs than 4th-CBs during RIPV ablations (15/41[36.6%] vs. 12/49[24.5%], p = 0.21). The TTI monitoring capability was significantly higher with 4th-CBs than 2nd-CBs (131/188[69.7%] vs. 83/160[51.9%] PVs, p = 0.0007). The difference was significant for right superior and left inferior PVs, but not for left superior PVs. Even if PVs requiring crossover were excluded, the total freeze duration (715±152 vs. 755±215 seconds, p = 0.31) tended to be shorter for 2nd-CBs than 4th-CBs. The incidence of phrenic nerve injury was similar for 2nd-CB and 4th-CB ablation (0/41 vs. 2/49, p = 0.12)

CONCLUSIONS:

The 4th-CB's shorter balloon tip enabled a significantly higher capability of TTI monitoring; however, it resulted in significantly lower rates of successful PVIs than the 2nd-CB, especially for the RIPVs.