Determination of ceftazidime in plasma by RP-HPLC and ultraviolet detection.
Biomed Chromatogr
; 35(7): e5104, 2021 Jul.
Article
em En
| MEDLINE
| ID: mdl-33629742
A simple high-performance liquid chromatography method for the determination of ceftazidime in plasma has been developed. Using an ultrafiltration technique samples were separated by reverse-phase high-performance liquid chromatography on a Symmetry C18 4.6 × 250 mm column (5.0 µm) and ultraviolet absorbance was measured at 260 nm. The mobile phase was a mixture of 10 mm potassium phosphate monobasic pH 2.5 with phosphoric acid and acetonitrile (90:10). The standard curve ranged from 0.1 to 100 µg/ml. Intra- and inter-assay variability for ceftazidime was <12%, and the average recovery was 89%. The lower limit of quantification was 0.1 µg/ml. This method has been used successfully to analyze frog plasma samples at this institution and it could be applied to other small volume samples in a clinical or research setting.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Espectrofotometria Ultravioleta
/
Ceftazidima
/
Cromatografia Líquida de Alta Pressão
/
Cromatografia de Fase Reversa
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
Limite:
Humans
Idioma:
En
Revista:
Biomed Chromatogr
Ano de publicação:
2021
Tipo de documento:
Article