ASSESSMENT OF INTEGRITY AND LEAD-EQUIVALENCE OF SHIELDED GARMENTS USING TWO-DIMENSIONAL X-RAY IMAGES FROM A COMPUTED TOMOGRAPHY SCANNER.
Kairn, T; Maxwell, S K; Trapp, J V; Crowe, S B.
Afiliação
  • Kairn T; Cancer Care Services, Royal Brisbane and Women's Hospital, Butterfield St, Herston 4029, Qld, Australia.
  • Maxwell SK; School of Chemistry and Physics, Queensland University of Technology, 2 George St, Brisbane 4000, Qld, Australia.
  • Trapp JV; Herston Biofabrication Institute, Metro North Hospital and Health Service, Butterfield St, Herston 4029, Qld, Australia.
  • Crowe SB; School of Information Technology and Electrical Engineering, University of Queensland, St Lucia, Qld 4072, Australia.
Radiat Prot Dosimetry ; 193(3-4): 155-164, 2021 May 17.
Article em En | MEDLINE | ID: mdl-33822208
Shielded garments are widely recommended for occupational radiation protection in diagnostic and interventional radiology. This study investigated a novel method for efficiently verifying shielded garment integrity while simultaneously acquiring data for lead-equivalence measurements, using two-dimensional topogram images from computed tomography (CT) scanners. This method was tested against more-conventional measurements with superficial and orthovoltage radiotherapy treatment beams, for 12 shielded garments containing 3 different lead-free shielding materials. Despite some energy-dependent results, all shielded garments approximately achieved their specified lead-equivalence for the energy range expected during clinical use for fluoroscopy procedures, except for three shielded skirts that required two layers of material to be overlapped at the front. All lead-equivalence measurements from CT topograms agreed with or conservatively underestimated the kV narrow-beam results. This method is potentially useful for independently assessing the shielding properties of new shielded garments and performing annual checks for damage or degradation of existing shielded garments.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Roupa de Proteção / Proteção Radiológica Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Roupa de Proteção / Proteção Radiológica Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2021 Tipo de documento: Article