Your browser doesn't support javascript.
loading
Radiation dose and diagnostic reference levels for four interventional radiology procedures: results of the prospective European multicenter survey EUCLID.
Schegerer, Alexander A; Frija, Guy; Paulo, Graciano; Jaschke, Werner; Tsapaki, Virginia; Repussard, Jacques; Damilakis, John.
Afiliação
  • Schegerer AA; Department of Radiation Protection and Image Processing Systems, Hirslanden AG, Hirslanden Corporate Office, Boulevard Lilienthal 2, 8152, Opfikon-Glattpark, Switzerland. alexander.schegerer@hirslanden.ch.
  • Frija G; Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany. alexander.schegerer@hirslanden.ch.
  • Paulo G; Paris Descartes University, Paris, France.
  • Jaschke W; ESTESC - Coimbra Health School, Medical Imaging and Radiotherapy Department, Instituto Politécnico de Coimbra, Rua 5 de Outubro, S. Martinho do Bispo, 3046-854, Coimbra, Portugal.
  • Tsapaki V; Department of Radiology, Medical University Innsbruck, Innsbruck, Austria.
  • Repussard J; Konstantopoulio General Hospital, Athens, Greece.
  • Damilakis J; , St Nom la Bretèche, France.
Eur Radiol ; 31(12): 9346-9360, 2021 Dec.
Article em En | MEDLINE | ID: mdl-33991223
ABSTRACT

OBJECTIVES:

To assess information reflecting radiation dose and define diagnostic reference levels (DRL) on a European basis for four interventional radiology (IR) procedures considering clinical indication, anatomical region, and procedure.

METHODS:

A prospective European study was performed to provide data on the IR procedures percutaneous recanalization of iliac arteries, percutaneous recanalization of femoropopliteal arteries, transarterial chemoembolization of hepatocellular carcinoma, and percutaneous transhepatic biliary drainage. Hospitals were asked to complete a questionnaire giving information on procedure, equipment, and protocol. Patient size and weight, experience of the operator graded in number of procedures performed, and complexity level of each procedure were reported. Sixteen hospitals from 13 countries could be surveyed. The percentiles of the kerma-area product, fluoroscopy time, cumulative air kerma at the interventional reference point, and number of images were determined. The impact of equipment, year of installation, and complexity level of the procedure on dose were analyzed.

RESULTS:

DRLs based on clinical indication were defined. Dose values varied considerably within hospitals, between them, and within each subgroup of complexity level. The use of state-of-the-art equipment reduced dose significantly by 52%. Although dose also varied within each subgroup of complexity level, for transarterial chemoembolization of hepatocellular carcinoma and percutaneous transhepatic biliary drainage, dose significantly correlated with complexity.

CONCLUSIONS:

This was the first study reporting exposure practice and defining DRLs based on clinical indication for four IR procedures on a European basis. These DRLs can serve as a baseline for comparison with local practice, the study as a guideline for future surveys. KEY POINTS • The use of state-of-the-art angiographic equipment reduces dose significantly. • A significant correlation between radiation dose and complexity level is found. • Dose values vary considerably, both within and between individual hospitals, and within each complexity level of interventional radiology procedure.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Quimioembolização Terapêutica / Carcinoma Hepatocelular / Neoplasias Hepáticas Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies / Qualitative_research Limite: Humans Idioma: En Revista: Eur Radiol Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Quimioembolização Terapêutica / Carcinoma Hepatocelular / Neoplasias Hepáticas Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies / Qualitative_research Limite: Humans Idioma: En Revista: Eur Radiol Ano de publicação: 2021 Tipo de documento: Article