Flow Dynamic Analysis by Contrast-Enhanced Imaging Techniques of Medium Cutoff Membrane Hemodialyzer.
Lorenzin, Anna; Golino, Gianlorenzo; de Cal, Massimo; Pajarin, Giordano; Savastano, Sergio; Lupi, Andrea; Sandini, Alessandra; Fiorin, Francesco; Ronco, Claudio.
Afiliação
  • Lorenzin A; International Renal Research Institute of Vicenza, Vicenza, Italy.
  • Golino G; Department of Nephrology, Dialysis and Transplantation, St. Bortolo Hospital, aULSS8 Berica, Vicenza, Italy.
  • de Cal M; International Renal Research Institute of Vicenza, Vicenza, Italy.
  • Pajarin G; Department of Medicine-DIMED, Section of Anesthesiology and Intensive Care Medicine, University of Padova, Padua, Italy.
  • Savastano S; International Renal Research Institute of Vicenza, Vicenza, Italy.
  • Lupi A; Department of Nephrology, Dialysis and Transplantation, St. Bortolo Hospital, aULSS8 Berica, Vicenza, Italy.
  • Sandini A; Department of Radiology, St. Bortolo Hospital, aULSS8 Berica, Vicenza, Italy.
  • Fiorin F; Department of Radiology, St. Bortolo Hospital, aULSS8 Berica, Vicenza, Italy.
  • Ronco C; Department of Nuclear Medicine, St. Bortolo Hospital, aULSS8 Berica, Vicenza, Italy.
Blood Purif ; 51(2): 138-146, 2022.
Article em En | MEDLINE | ID: mdl-34034259
INTRODUCTION: Medium cutoff (MCO) membranes represent an interesting innovation in the field of hemodialysis. Given the correlation between large (PM >25 kDa) middle molecules (LMM) and clinical outcomes, the possibility to broaden the spectrum of solutes removed in hemodialysis with MCO membranes introduces a new perspective for end-stage kidney disease patients. Due to low diffusion coefficients of LMM, the use of convection is required to maximize extracorporeal clearance. High convective rates are achieved with high-flux membranes in hemodiafiltration, a technique not available in the US. In case of the MCO membrane, remarkable clearances of LMM are achieved combining the permeability of the membrane with a significant amount of internal convection. The mechanism of filtration-backfiltration inside the dialyzer enables effective removal of LMM in a technique called expanded hemodialysis (HDx). Given such theoretical explanation, it is important to demonstrate the blood and ultrafiltration rheology inside the MCO dialyzer. METHOD: This study for the first time describes flow dynamic parameters and internal cross-filtration, thanks to specific radiology and nuclear imaging techniques. RESULTS: Flow dynamic analysis of the blood and dialysate compartment confirms excellent distribution of velocities and an excellent matching of blood and dialysate. Average blood flow velocity allows for wall shear rates adequate to avoid protein stagnation at the blood membrane interface and increase in blood viscosity. Cross-filtration analysis demonstrates a remarkable filtration/backfiltration flux reaching values >30 mL/min at a blood flow of 300 mL/min and zero net filtration. CONCLUSION: The MCO dialyzer Theranova 400 appears to have a design optimized to perform expanded hemodialysis (HDx).