The Utility of iCare HOME Tonometry for Detection of Therapy-Related Intraocular Pressure Changes in Glaucoma and Ocular Hypertension.
Scott, Anthony T; Kanaster, Karen; Kaizer, Alexander M; Young, Cara Capitena; Pantcheva, Mina B; Ertel, Monica K; Kahook, Malik Y; Seibold, Leonard K.
Afiliação
  • Scott AT; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Kanaster K; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Kaizer AM; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Young CC; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Pantcheva MB; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Ertel MK; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Kahook MY; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado.
  • Seibold LK; Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado. Electronic address: leonard.seibold@cuanschutz.edu.
Ophthalmol Glaucoma ; 5(1): 85-93, 2022.
Article em En | MEDLINE | ID: mdl-34082179
ABSTRACT

PURPOSE:

To assess whether iCare HOME rebound tonometry can detect therapy-related changes during self-monitoring of intraocular pressure (IOP).

DESIGN:

Prospective clinical trial.

PARTICIPANTS:

A total of 43 eyes (n = 27 subjects) with open-angle glaucoma or ocular hypertension were enrolled during standard-of-care clinic visits. Participants were grouped into control eyes managed on stable therapy (n = 18 eyes) or therapy change eyes undergoing selective laser trabeculoplasty (SLT, n = 8 eyes), initiating topical therapy (n = 8 eyes), or adding a second medication to existing monotherapy (n = 9 eyes).

METHODS:

Subjects recorded IOP 4 times daily for 1 week using iCare HOME tonometry. Upon tonometer return, subjects underwent SLT or new medication start; an additional week of iCare HOME measurements was collected after 4 to 6 weeks. Control subjects recorded an additional week of measurements after 6 weeks. Measurements were grouped into 4 time periods (5-10 am, 10 am to 3 pm, 3-8 pm, 8 pm to 1 am). Goldmann applanation tonometry (GAT) was performed at each study visit for comparison. MAIN OUTCOME

MEASURES:

Detection of therapy response defined as an IOP reduction of ≥20%.

RESULTS:

For eyes that demonstrated a therapy response by GAT (n = 11), iCare HOME detected a therapy response in 90.9% of eyes in ≥1 time period and 45.5% of eyes in all 4 time periods. In eyes without a GAT-measured therapy response (n = 14), iCare HOME detected a response for 71.4% (n = 10) of eyes in ≥1 time period and for 7.1% of eyes (n = 1) at all 4 time periods. In treatment eyes, intraday and interday average minimum and maximum IOP, as well as interday IOP range, were significantly reduced after therapy without a significant change in intraday IOP range. Control group eyes did not demonstrate a significant change in average IOP minimum, maximum, or range between study weeks.

CONCLUSIONS:

Home tonometry with iCare HOME reliably detects therapy-related IOP changes in patients with glaucoma and ocular hypertension. Treatment responses correlated well with in-office GAT and may detect treatment responses missed by GAT. Intraocular pressure measurements via home tonometry provide additional clinical information regarding intraday and interday IOP fluctuation beyond standard of care in office GAT measurements. The iCare HOME is a valuable tool to monitor therapeutic efficacy in patients with glaucoma.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glaucoma / Glaucoma de Ângulo Aberto / Hipertensão Ocular Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies Limite: Humans Idioma: En Revista: Ophthalmol Glaucoma Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glaucoma / Glaucoma de Ângulo Aberto / Hipertensão Ocular Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies Limite: Humans Idioma: En Revista: Ophthalmol Glaucoma Ano de publicação: 2022 Tipo de documento: Article