Validation of the BPro radial pulse waveform acquisition device in pregnancy and gestational hypertensive disorders.
Blood Press Monit
; 26(5): 380-384, 2021 Oct 01.
Article
em En
| MEDLINE
| ID: mdl-34128489
ABSTRACT
OBJECTIVE:
To validate the BPro blood pressure (BP) wrist device for use in pregnancy and preeclampsia according to the Universal Standard protocol. PARTICIPANTS ANDMETHODS:
BP was measured sequentially in 45 pregnant women (including 15 with preeclampsia, 15 with gestational hypertension and 15 who remained normotensive) alternating between a mercury sphygmomanometer and BPro device.RESULTS:
The BPro is accurate in pregnancy with a mean device-observer difference of -1.7 ± 6.1 and 0.1 ± 4.6 mmHg for SBP and DBP, respectively. In women with preeclampsia, BPro also met the validation criteria for the Universal Standard protocol with a mean device-observer difference of -2.7 ± 7.1 and 0.3 ± 4.7 mmHg for SBP and DBP, respectively. However, the number of absolute BP differences within 5 mmHg was considerably fewer in those with preeclampsia when compared to the other two subgroups.CONCLUSION:
The BPro device can be recommended for BP measurement in pregnancy but should be used with caution in those with confirmed preeclampsia.