Your browser doesn't support javascript.
loading
Test-Retest Reliability of Functional Electromechanical Dynamometer on Five Sit-to-Stand Measures in Healthy Young Adults.
Jerez-Mayorga, Daniel; Huerta-Ojeda, Álvaro; Chirosa-Ríos, Luis Javier; Guede-Rojas, Francisco; Guzmán-Guzmán, Iris Paola; Intelangelo, Leonardo; Miranda-Fuentes, Claudia; Delgado-Floody, Pedro.
Afiliação
  • Jerez-Mayorga D; Faculty of Rehabilitation Sciences, Universidad Andres Bello, Santiago 7591538, Chile.
  • Huerta-Ojeda Á; Grupo de Investigación en Salud, Actividad Física y Deporte ISAFYD, Universidad de Las Américas, Sede Viña del Mar 2531098, Chile.
  • Chirosa-Ríos LJ; Department Physical Education and Sports, Faculty of Sport Sciences, University of Granada, 18011 Granada, Spain.
  • Guede-Rojas F; Faculty of Rehabilitation Sciences, Universidad Andres Bello, Concepción 4030000, Chile.
  • Guzmán-Guzmán IP; Faculty of Chemical-Biological Sciences, Universidad Autόnoma de Guerrero, Chilpancingo 39087, Mexico.
  • Intelangelo L; Musculoskeletal Research Group, University Center for Assistance, Teaching and Research, University of Gran Rosario, Rosario 2000, Argentina.
  • Miranda-Fuentes C; Faculty of Rehabilitation Sciences, Universidad Andres Bello, Santiago 7591538, Chile.
  • Delgado-Floody P; Department of Physical Education, Sports and Recreation, Universidad de La Frontera, Temuco 4811230, Chile.
Article em En | MEDLINE | ID: mdl-34202138
ABSTRACT

Background:

The purpose of this study was to determine the reliability for the strength and movement velocity of the concentric phase from the five Sit-to-Stand (5STS), using three incremental loads measured by a functional electromechanical dynamometer (FEMD) in healthy young adults.

Methods:

The average and peak strength and velocity values of sixteen healthy adults (mean ± standard deviation (SD) age = 22.81 ± 2.13 years) were recorded at 5, 10 and 15 kg. To evaluate the reliability of FEMD, the intraclass correlation coefficient (ICC), standard error of measurement (SEM) and coefficient of variation (CV) were obtained.

Results:

Reliability was high for the 10 kg (CV range 3.70-4.18%, ICC range 0.95-0.98) and 15 kg conditions (CV range 1.64-3.02%, ICC 0.99) at average and peak strength, and reliability was high for the 5 kg (CV range 1.71-2.84%, ICC range 0.96-0.99), 10 kg (CV range 0.74-1.84%, ICC range 0.99-1.00) and 15 kg conditions (CV range 0.79-3.11%, ICC range 0.99-1.00) at average and peak velocity.

Conclusions:

The findings of this study demonstrate that FEMD is a reliable instrument to measure the average and peak strength and velocity values during the five STS in healthy young adults.