Your browser doesn't support javascript.
loading
The feasibility of an approximate irregular field dose distribution simulation program applied to a respiratory motion compensation system.
Ting, Lai-Lei; Liao, Ai-Ho; Ganesan, Muthusankar; Kuo, Chia-Chun; Yu, Hsiao-Wei; Chen, Po-Jung; Jeng, Shiu-Chen; Chiou, Jeng-Fong; Chuang, Ho-Chiao.
Afiliação
  • Ting LL; Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan.
  • Liao AH; Graduate Institute of Biomedical Engineering, National Taiwan University of Science and Technology, Taipei, Taiwan; Department of Biomedical Engineering, National Defense Medical Center, Taipei, Taiwan.
  • Ganesan M; Department of Mechanical Engineering, National Taipei University of Technology, Taipei, Taiwan; Department of Industrial Chemistry, Alagappa University, Karaikudi 630001, Tamil Nadu, India.
  • Kuo CC; Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan; Department of Radiation Oncology, Wan Fang Hospital, Taipei Medical University, Taipei, Taiwan; School of Health Care Administration, College of Management, Taipei Medical University, Taipei, Taiwan.
  • Yu HW; Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan; School of Biomedical Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei, Taiwan.
  • Chen PJ; Department of Mechanical Engineering, National Taipei University of Technology, Taipei, Taiwan.
  • Jeng SC; Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan; School of Dentistry, College of Oral Medicine, Taipei Medical University, Taipei, Taiwan.
  • Chiou JF; Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan; Department of Radiology, School of Medicine, College of Medicine, Taipei Medical University, Taipei, Taiwan; Taipei Cancer Center, Taipei Medical University, Taipei, Taiwan.
  • Chuang HC; Department of Mechanical Engineering, National Taipei University of Technology, Taipei, Taiwan. Electronic address: hchuang@mail.ntut.edu.tw.
Phys Med ; 88: 117-126, 2021 Aug.
Article em En | MEDLINE | ID: mdl-34237677
ABSTRACT

PURPOSE:

This study optimized our previously proposed simulation program for the approximate irregular field dose distribution (SPAD) and applied it to a respiratory motion compensation system (RMCS) and respiratory motion simulation system (RMSS). The main purpose was to rapidly analyze the two-dimensional dose distribution and evaluate the compensation effect of the RMCS during radiotherapy.

METHODS:

This study modified the SPAD to improve the rapid analysis of the dose distribution. In the experimental setup, four different respiratory signal patterns were input to the RMSS for actuation, and an ultrasound image tracking algorithm was used to capture the real-time respiratory displacement, which was input to the RMCS for actuation. A linear accelerator simultaneously irradiated the EBT3 film. The gamma passing rate was used to verify the dose similarity between the EBT3 film and the SPAD, and conformity index (CI) and compensation rate (CR) were used to quantify the compensation effect.

RESULTS:

The Gamma passing rates were 70.48-81.39% (2%/2mm) and 88.23-96.23% (5%/3mm) for various collimator opening patterns. However, the passing rates of the SPAD and EBT3 film ranged from 61.85% to 99.85% at each treatment time point. Under the four different respiratory signal patterns, CR ranged between 21% and 75%. After compensation, the CI for 85%, 90%, and 95% isodose constraints were 0.78, 0.57, and 0.12, respectively.

CONCLUSIONS:

This study has demonstrated that the dose change during each stage of the treatment process can be analyzed rapidly using the improved SPAD. After compensation, applying the RMCS can reduce the treatment errors caused by respiratory movements.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Algoritmos Aspecto: Implementation_research Idioma: En Revista: Phys Med Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Algoritmos Aspecto: Implementation_research Idioma: En Revista: Phys Med Ano de publicação: 2021 Tipo de documento: Article