[Application value of liquid crystal digital display goniometer in total hip arthroplasty].
Hou, Jing-Zhao; Wang, Hai-Hong; Cheng, Yan-Xiao; Bao, Hong-Wei; Wang, Yu-Ji; Sun, Yu.
Afiliação
  • Hou JZ; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
  • Wang HH; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
  • Cheng YX; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
  • Bao HW; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
  • Wang YJ; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
  • Sun Y; Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China.
Zhongguo Gu Shang ; 34(7): 597-600, 2021 Jul 25.
Article em Zh | MEDLINE | ID: mdl-34318632
ABSTRACT

OBJECTIVE:

To investigate the application value of liquid crystal digital display goniometer in total hip arthroplasty.

METHODS:

From January 2018 to December 2019, 83 patients underwent primary total hip arthroplasty, including 28 males and 55 females, aged 42 to 81 (70.4±7.9) years. There were 63 cases of femoral neck fracture and 20 cases of avascular necrosis of femoral head. All patients used liquid crystal digital goniometer to control the anteversion of acetabular cup prosthesis during operation, and CT scanning was used to measure the anteversion of acetabular cup after operation. The two methods were compared to understand the accuracy of using liquid crystal digital goniometer.

RESULTS:

Postoperative CT measurement showed that the acetabular anteversion of all patients was in the safe area advocated by Lewinnek. The anteversion angle of acetabular cup measured by liquid crystal digital goniometer was 14.20(12.80 to 15.40)°, and the anteversion angle of acetabular cup measured by postoperative CT scan was 14.20 (13.40 to 15.50)°. There was no significant difference between the two (Z=-1.725, P=0.085).

CONCLUSION:

It is an accurate and reliable method to control the anteversion of acetabular cup with liquid crystal digital display angle instrument, which has a good auxiliary reference value.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Cristais Líquidos / Prótese de Quadril Tipo de estudo: Observational_studies Limite: Female / Humans / Male Idioma: Zh Revista: Zhongguo Gu Shang Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Cristais Líquidos / Prótese de Quadril Tipo de estudo: Observational_studies Limite: Female / Humans / Male Idioma: Zh Revista: Zhongguo Gu Shang Ano de publicação: 2021 Tipo de documento: Article