Air leaks: Stapling affects porcine lungs biomechanics.
Bonnet, Bénédicte; Tabiai, Ilyass; Rakovich, George; Gosselin, Frédérick P; Villemure, Isabelle.
Afiliação
  • Bonnet B; Department of Mechanical Engineering, Polytechnique Montréal, 2500 Chemin de Polytechnique, Montreal, QC, H3T 1J4, Canada. Electronic address: benedicte.bonnet@polymtl.ca.
  • Tabiai I; Department of Mechanical Engineering, Polytechnique Montréal, 2500 Chemin de Polytechnique, Montreal, QC, H3T 1J4, Canada. Electronic address: ilyass.tabiai@polymtl.ca.
  • Rakovich G; Department of Thoracic Surgery, University of Montréal, 2900 Boulevard Edouard-Montpetit, Montreal, QC, H3T 1J4, Canada. Electronic address: george.rakovich@umontreal.ca.
  • Gosselin FP; Department of Mechanical Engineering, Polytechnique Montréal, 2500 Chemin de Polytechnique, Montreal, QC, H3T 1J4, Canada. Electronic address: frederick.gosselin@polymtl.ca.
  • Villemure I; Department of Mechanical Engineering, Polytechnique Montréal, 2500 Chemin de Polytechnique, Montreal, QC, H3T 1J4, Canada. Electronic address: isabelle.villemure@polymtl.ca.
J Mech Behav Biomed Mater ; 125: 104883, 2022 01.
Article em En | MEDLINE | ID: mdl-34678619
During thoracic operations, surgical staplers resect cancerous tumors and seal the spared lung. However, post-operative air leaks are undesirable clinical consequences: staple legs wound lung tissue. Subsequent to this trauma, air leaks from lung tissue into the pleural space. This affects the lung's physiology and patients' recovery. The objective is to biomechanically and visually characterize porcine lung tissue with and without staples in order to gain knowledge on air leakage following pulmonary resection. Therefore, a syringe pump filled with air inflates and deflates eleven porcine lungs cyclically without exceeding 10 cmH2O of pressure. Cameras capture stereo-images of the deformed lung surface at regular intervals while a microcontroller simultaneously records the alveolar pressure and the volume of air pumped. The raw images are then used to compute tri-dimensional displacements and strains with the Digital Image Correlation method (DIC). Air bubbles originated at staple holes of inner row from exposed porcine lung tissue due to torn pleural on costal surface. Compared during inflation, left upper or lower lobe resections have similar compliance (slope of the pressure vs volume curve), which are 9% lower than healthy lung compliance. However, lower lobes statistically burst at lower pressures than upper lobes (p-value<0.046) in ex vivo conditions confirming previous clinical in vivo studies. In parallel, the lung deformed mostly in the vicinity of staple holes and presented maximum shear strain near the observed leak location. To conclude, a novel technique DIC provided concrete evidence of the post-operative air leaks biomechanics. Further studies could investigate causal relationships between the mechanical parameters and the development of an air leak.