Your browser doesn't support javascript.
loading
Reverse RAMP (REVRAMP) pacing: A novel anti tachycardia pacing technique.
Tayebjee, Muzahir Hassan; Bowes, Robert; Stegemann, Berthold; Holden, Arun V.
Afiliação
  • Tayebjee MH; Department of Cardiology, Leeds General Infirmary, Leeds, UK. Electronic address: muzahir.tayebjee@nhs.net.
  • Bowes R; Department of Cardiology, Leeds General Infirmary, Leeds, UK.
  • Stegemann B; University of Aston, Birmingham, UK.
  • Holden AV; Faculty of Biological Sciences, University of Leeds, Leeds, UK.
Indian Pacing Electrophysiol J ; 22(1): 54-57, 2022.
Article em En | MEDLINE | ID: mdl-34775047
ABSTRACT
Anti-tachycardia pacing (ATP) is frequently used to terminate ventricular tachycardia (VT), however it is not always successful and may accelerate VT requiring defibrillation. REVRAMP is a novel concept of ATP that involves delivering pacing at a faster rate than VT, but instead of abruptly terminating pacing after eight beats, pacing is gradually slowed until VT continues or normal rhythm is restored. In a pilot study we show that REVRAMP can restore normal rhythm, and that if REVRAMP is unsuccessful, VT is not accelerated.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Indian Pacing Electrophysiol J Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Indian Pacing Electrophysiol J Ano de publicação: 2022 Tipo de documento: Article