[The Method for Estimation of Radiation Dose of Eye Lens Covered by Radiation Protection Glasses Using Personal Dosimeter for Trunk of the Body and X-ray Shielding Material].
Osanai, Minoru; Sato, Kana; Sato, Hidenori; Kudo, Kohsei; Hosokawa, Shota; Fujita, Akira; Kitajima, Maiko; Tsushima, Megumi; Komiya, Nobuhiro; Kudo, Masaya; Saito, Yoko.
Afiliação
  • Osanai M; Hirosaki University Graduate School of Health Sciences.
  • Sato K; Hirosaki University School of Health Sciences (Current address: Division of Radiology, Towada City Hospital).
  • Sato H; Hirosaki University School of Health Sciences (Current address: Division of Radiology, Saiseikai Yokohamashi Tobu Hospital).
  • Kudo K; Hirosaki University Graduate School of Health Sciences.
  • Hosokawa S; Hirosaki University Graduate School of Health Sciences.
  • Fujita A; Division of Radiology, Narumi Hospital.
  • Kitajima M; Hirosaki University Graduate School of Health Sciences.
  • Tsushima M; Hirosaki University Graduate School of Health Sciences.
  • Komiya N; Hirosaki University Graduate School of Health Sciences (Current address: Department of Radiology, National Hospital Organization Hirosaki National Hospital).
  • Kudo M; Hirosaki University Graduate School of Health Sciences (Current address: Department of Radiology, Southern TOHOKU Proton Therapy Center, Southern TOHOKU Research Institute for Neuroscience).
  • Saito Y; Hirosaki University Graduate School of Health Sciences.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 77(12): 1432-1443, 2021.
Article em Ja | MEDLINE | ID: mdl-34924480
PURPOSE: Radiation protection glasses is a useful tool for dose reduction of eye lens. However, the method to evaluate the dose of eye lens, which is covered with radiation protection glasses in actual radiation management situation, has not been established. We invented the easy-to-use procedure to estimate the dose of eye lens inside of radiation protection glasses for occupationally exposed person using an X-ray shielding material and an additional personal dosimeter for trunk of the body. The objective of this study was to evaluate the feasibility of the new method. METHODS: The radiation field in interventional radiology (IR) was reconstructed. A personal dosimeter was set to the eye position and neck position of medical staff phantom. Each dosimeter was covered by radiation protection glasses or X-ray shielding material, and the 1 cm or 70 mm dose equivalent quantity in the neck position was compared to the 3 mm dose equivalent quantity in the eye position. RESULTS: The dose equivalent quantity measured inside of X-ray shielding material in the neck position was similar to that of inside radiation protection glasses. Thus, the X-ray attenuation by radiation protection glasses was able to be simulated using the X-ray shielding material. CONCLUSION: It was suggested that the dose of eye lens inside of radiation protection glasses could be easily estimated by covering the conventional personal dosimeter for trunk of the body with the X-ray shielding material.