Your browser doesn't support javascript.
loading
Dual-layer spectral detector CT arthrography of the shoulder: assessment of image quality and value in differentiating calcium from iodine.
Yoon, Hyeyoung; Kang, Yusuhn; Kim, Hyo Jin; Lee, Eugene; Ahn, Joong Mo; Lee, Joon Woo.
Afiliação
  • Yoon H; Department of Radiology, 65462Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
  • Kang Y; Department of Radiology, 65462Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
  • Kim HJ; Department of Radiology, 65462Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
  • Lee E; Department of Radiology, Seoul Metropolitan Government - Seoul National University Boramae Medical Center, 65633Seoul National University College of Medicine, Seoul, Republic of Korea.
  • Ahn JM; Department of Radiology, 65462Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
  • Lee JW; Department of Radiology, 65462Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
Acta Radiol ; 64(2): 638-647, 2023 Feb.
Article em En | MEDLINE | ID: mdl-35300534
ABSTRACT

BACKGROUND:

Dual-layer spectral detector computed tomography (DLCT) may potentially improve CT arthrography through enhanced image quality and analysis of the chemical composition of tissue.

PURPOSE:

To evaluate the image quality of monoenergetic reconstructions from DLCT arthrography of the shoulder and assess the additional diagnostic value in differentiating calcium from iodine. MATERIAL AND

METHODS:

Images from consecutive shoulder DLCT arthrography examinations performed between December 2016 and February 2018 were retrospectively reviewed for hyperattenuating lesions within the labrum and tendons. The mean attenuation of the target lesion, noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) of the virtual monoenergetic images obtained at 40-200 keV were compared with conventional 140-kVp images. Two evaluators independently classified each target lesion as contrast media or calcification, without and with DLCT spectral data. Receiver operating curve (ROC) analysis was performed to assess the diagnostic performance of shoulder DLCT arthrography, without and with the aid of spectral data.

RESULTS:

The study included 20 target lesions (18 DLCT arthrography examinations of 17 patients). The SNRs of the monoenergetic images at 40-60 keV were significantly higher than those of conventional images (P < 0.05). The CNRs of the monoenergetic images at 40-70 keV were significantly higher than those of conventional images (P < 0.001). The ability to differentiate calcium from iodine, without and with DLCT spectral data, did not significantly differ (P = 0.441 and P = 0.257 for reviewers 1 and 2, respectively).

CONCLUSION:

DLCT had no additive value in differentiating calcium from iodine in small, hyperattenuating lesions in the labrum and tendons.