Your browser doesn't support javascript.
loading
Stability and in vivo safety of gold, titanium nitride and parylene C coatings on NdFeB magnets implanted in muscles towards a new generation of myokinetic prosthetic limbs.
Iacovacci, Veronica; Naselli, Irene; Salgarella, Alice Rita; Clemente, Francesco; Ricotti, Leonardo; Cipriani, Christian.
Afiliação
  • Iacovacci V; The BioRobotics Institute, Scuola Superiore Sant'Anna Piazza Martiri della Libertà 33 56127 Pisa Italy christian.cipriani@santannapisa.it.
  • Naselli I; Department of Excellence in Robotics & AI, Scuola Superiore Sant'Anna Piazza Martiri della Libertà 33 56127 Pisa Italy.
  • Salgarella AR; The BioRobotics Institute, Scuola Superiore Sant'Anna Piazza Martiri della Libertà 33 56127 Pisa Italy christian.cipriani@santannapisa.it.
  • Clemente F; The BioRobotics Institute, Scuola Superiore Sant'Anna Piazza Martiri della Libertà 33 56127 Pisa Italy christian.cipriani@santannapisa.it.
  • Ricotti L; The BioRobotics Institute, Scuola Superiore Sant'Anna Piazza Martiri della Libertà 33 56127 Pisa Italy christian.cipriani@santannapisa.it.
  • Cipriani C; Prensilia SRL Viale Rinaldo Piaggio 32 56025 Pontedera Italy.
RSC Adv ; 11(12): 6766-6775, 2021 Feb 04.
Article em En | MEDLINE | ID: mdl-35423178
ABSTRACT
Rare earth magnets are the elective choice when high magnetic field density is required and they are particularly intriguing for inclusion in implantable devices. A safe implantation of NdFeB magnets in muscles would enable the control of limb prostheses using a myokinetic interface i.e., direct control of artificial limb movements by means of magnetic tracking of residual muscle contractions. However, myokinetic prosthesis control is prevented by NdFeB magnets poor biocompatibility, at present. Here we investigated three biocompatible materials as NdFeB magnet coating candidates, namely gold, titanium nitride and parylene C, which have not been analyzed in a systematic way for this purpose, so far. In vitro testing in a tissue-mimicking environment and upon contact with C2C12 myoblasts enabled assessment of the superiority of parylene C coated magnets in terms of corrosion prevention and lack of cytotoxicity. In addition, parylene C coated magnets implanted in rabbit muscles for 28 days confirmed, both locally and systemically, their biocompatibility, with a lack of irritation and toxicity associated with the implant. These findings pave the way towards the development of implantable devices based on permanent magnets and of a new generation of limb prostheses.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: RSC Adv Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: RSC Adv Ano de publicação: 2021 Tipo de documento: Article