Your browser doesn't support javascript.
loading
Outcomes of XEN 63 Device at 18-Month Follow-Up in Glaucoma Patients: A Two-Center Retrospective Study.
Fea, Antonio Maria; Menchini, Martina; Rossi, Alessandro; Posarelli, Chiara; Malinverni, Lorenza; Figus, Michele.
Afiliação
  • Fea AM; Struttura Complessa Oculistica, Città Della Salute e Della Scienza di Torino, Dipartimento di Scienze Chirurgiche-Università degli Studi di Torino, 10126 Torino, Italy.
  • Menchini M; Department of Surgical, Medical and Molecular Pathology and Critical Care Medicine-University of Pisa, 56126 Pisa, Italy.
  • Rossi A; Struttura Complessa Oculistica, Città Della Salute e Della Scienza di Torino, Dipartimento di Scienze Chirurgiche-Università degli Studi di Torino, 10126 Torino, Italy.
  • Posarelli C; Department of Surgical, Medical and Molecular Pathology and Critical Care Medicine-University of Pisa, 56126 Pisa, Italy.
  • Malinverni L; Struttura Complessa Oculistica, Città Della Salute e Della Scienza di Torino, Dipartimento di Scienze Chirurgiche-Università degli Studi di Torino, 10126 Torino, Italy.
  • Figus M; Department of Surgical, Medical and Molecular Pathology and Critical Care Medicine-University of Pisa, 56126 Pisa, Italy.
J Clin Med ; 11(13)2022 Jun 30.
Article em En | MEDLINE | ID: mdl-35807088
ABSTRACT

Background:

Glaucoma surgery has significantly evolved over the last years. This paper aims to evaluate the midterm clinical outcomes of the XEN63 device in a real-world scenario.

Methods:

A retrospective clinical study was conducted on consecutive patients who underwent an XEN63 implant insertion, either alone or in combination with phacoemulsification. The primary endpoint was the mean intraocular pressure (IOP) at the month 18 visit.

Results:

Twenty-three eyes (23 patients) were included in the analysis. The preoperative IOP was significantly lowered from 27.0 ± 7.8 mmHg to 14.1 ± 3.4 mmHg at month 18 (p < 0.0001). At month 18, 14 (77.8%) and 11 (61.1%) eyes had an IOP ≤ 16 mmHg and ≤ 14 mmHg, respectively, without ocular hypotensive medication. The mean number of ocular hypotensive medication taken was significantly reduced from 2.3 ± 0.9 drugs at baseline to 1.0 ± 1.4 drugs at month 18 (p = 0.0020). Four (17.4%) eyes had hypotony at postoperative day 1, which was successfully resolved without sequelae. Four (17.4%) eyes underwent a needling procedure and four (17.4%) eyes underwent additional surgeries.

Conclusions:

The XEN63, either alone or in combination with cataract surgery, significantly lowered the IOP and reduced the need for ocular hypotensive drugs over a period of 18 months.