Dose Optimization for Single-Channel Vaginal Cylinder High-Dose-Rate Brachytherapy: A Double Prescription Method for Patients With Endometrial Adenocarcinoma.
Wu, Xingen; McDonald, Andrew; Shen, Sui; Stanley, Dennis; Popple, Richard; Marcrom, Samuel; Kim, Robert.
Afiliação
  • Wu X; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • McDonald A; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • Shen S; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • Stanley D; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • Popple R; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • Marcrom S; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
  • Kim R; Radiation Oncology, University of Alabama at Birmingham Heersink School of Medicine, Birmingham, USA.
Cureus ; 14(6): e26303, 2022 Jun.
Article em En | MEDLINE | ID: mdl-35911294
Purpose This study aimed to explore the relationship between applicator surface dose and 5 mm-depth dose and to optimize both locations simultaneously for three most used cylinder sizes (2.5, 3.0, and 3.5 cm in diameter) in treating patients with endometrial adenocarcinoma. Materials and methods A total of 216 plans were created for each dose level and applicator size. For each dose level, four plans were created with single or double prescription doses. For plans with double prescription doses, the dose constraints were applied to all those points on the surface and 5 mm depth and optimize the two sites simultaneously.  Results A dose table between surface and 5 mm depth and its fifth order polynomial mapping functions were established for each applicator size, so any prescribed dose at one site can find the prescription dose on the other site in optimization on both locations. For plans with a 5 mm-depth prescription, the maximum dose on the surface can be reduced from 145% to 133% if the surface prescription dose is also used; for plans with surface dose prescription, the minimum dose and mean dose can be improved by 2% if 5 mm-depth dose prescription is also used in optimization. Conclusion Dose table and their mapping functions between surface prescription dose and their corresponding 5 mm-depth doses were created. A new optimization method that uses two prescription doses on both surface and 5 mm-depth sites was proposed to reduce the hot dose on the surface and improve the cold dose at 5 mm depth.