Ladder-Like Low Iodine Delivery Rate Injection Protocols Based on BMI at 70 KV by Automated Tube Voltage Selection in Coronary CTA.
Contrast Media Mol Imaging
; 2022: 7804015, 2022.
Article
em En
| MEDLINE
| ID: mdl-35924071
ABSTRACT
Objective:
To evaluate the feasibility of reducing the injection velocity and volume of contrast agent according to BMI, and the effect of body weight (BW), body surface areaï¼BSAï¼, body mass indexï¼BMIï¼ï¼and blood volume (BV) on aortic contrast enhancement when the voltage of third-generation dual-source CT is selected at 70 KV.Methods:
A total of 280 patients selected at 70 KV were randomly divided into an experimental group and a control group. Each group was divided into 7 subgroups according to BMI ≤20, 20-21, 21-22, 22-23, 23-24, 24-25, and 25-26. The experimental group uses 2.3/2.4/2.5/2.6/2.7/2.8/2.9 ml/s injection speed with 350 mgI/ml contrast agents according to the subgroups; injection time was fixed at 10 s. In the control group, the fixed injection flow rate was 3.5 ml/s, time was 12 s with a total of 42 ml. Subjects in both groups were inspected to adaptive prospective ECG-gating sequence scanning, and subjective and objective image quality of the two groups were compared using Student's t-test. BMI, BSA, and BV were calculated from the patient's body weight and height. We assess the relationship between aortic attenuation and BW, BMI, BV, and BSA using regression analysis or correlation analysis.Results:
Significant differences exist in vascular enhancement between the two groups; SNR and CNR of objective image quality in the experimental group were lower than those in the control group (P < 0.05). Both groups had the same subjective image scores (P > 0.05). The number of vessels in the optimal enhancement range counts more in the experimental group than in the control group (χ2 value = 334.25, P ï¼ 0.05). In the control group, a weak to medium correlation was seen between vascular enhancement and BMI (r = -0.20), BW (r = -0.42), BSA (r = -0.46), and BV (r = -0.48) (P ï¼ 0.05 for all).Conclusions:
Compared to BW, BSA, and BV, a weaker negative correlation exists between vascular enhancement and BMI when ATVS selects 70 KV. However, as a much easier way to operate, the stepped low flow and low-contrast agent injection based on BMI was feasible, and the image quality was more homogenized than that of the control group.