Your browser doesn't support javascript.
loading
Neutral-Coating Capillary Electrophoresis Coupled with High-Resolution Mass Spectrometry for Top-Down Identification of Hemoglobin Variants.
Luo, Ruben Yiqi; Wong, Carolyn; Xia, James Qiangwei; Glader, Bertil E; Shi, Run-Zhang; Zehnder, James L.
Afiliação
  • Luo RY; Department of Pathology, School of Medicine, Stanford University, Stanford, CA, USA.
  • Wong C; Clinical Laboratories, Stanford Health Care, Palo Alto, CA, USA.
  • Xia JQ; Clinical Laboratories, Stanford Health Care, Palo Alto, CA, USA.
  • Glader BE; CMP Scientific Corp, Brooklyn, NY, USA.
  • Shi RZ; Department of Pathology, School of Medicine, Stanford University, Stanford, CA, USA.
  • Zehnder JL; Department of Pediatrics, School of Medicine, Stanford University, Stanford, CA, USA.
Clin Chem ; 69(1): 56-67, 2023 01 04.
Article em En | MEDLINE | ID: mdl-36308334
ABSTRACT

BACKGROUND:

Identification of hemoglobin (Hb) variants is of significant value in the clinical diagnosis of hemoglobinopathy. However, conventional methods for identification of Hb variants in clinical laboratories can be inadequate due to the lack of structural characterization. We describe the use of neutral-coating capillary electrophoresis coupled with high-resolution mass spectrometry (CE-HR-MS) to achieve high-performance top-down identification of Hb variants.

METHODS:

An Orbitrap Q-Exactive Plus mass spectrometer was coupled with an ECE-001 capillary electrophoresis (CE) unit through an EMASS-II ion source. A PS1 neutral-coating capillary was used for CE. Samples of red blood cells were lysed in water and diluted in 10 mM ammonium formate buffer for analysis. Deconvolution of raw mass spectrometry data was carried out to merge multiple charge states and isotopic peaks of an analyte to obtain its monoisotopic mass.

RESULTS:

The neutral-coating CE could baseline separate individual Hb subunits dissociated from intact Hb forms, and the HR-MS could achieve both intact-protein analysis and top-down analysis of analytes. A number of patient samples that contain Hb subunit variants were analyzed, and the variants were successfully identified using the CE-HR-MS method.

CONCLUSIONS:

The CE-HR-MS method has been demonstrated as a useful tool for top-down identification of Hb variants. With the ability to characterize the primary structures of Hb subunits, the CE-HR-MS method has significant advantages to complement or partially replace the conventional methods for the identification of Hb variants.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletroforese Capilar / Hemoglobinopatias Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Clin Chem Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletroforese Capilar / Hemoglobinopatias Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Clin Chem Ano de publicação: 2023 Tipo de documento: Article