Failure Mode and Effects Analysis for Experimental Use of FLASH on a Clinical Accelerator.
Rahman, Mahbubur; Zhang, Rongxiao; Gladstone, David J; Williams, Benjamin B; Chen, Erli; Dexter, Chad A; Thompson, Lawrence; Bruza, Petr; Pogue, Brian W.
Afiliação
  • Rahman M; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; University of Texas Southwestern Medical Center, Dallas, Texas. Electronic address: Mahbubur.Rahman.th@dartmouth.edu.
  • Zhang R; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; Department of Medicine, Radiation Oncology, Geisel School of Medicine, Dartmouth College Hanover, New Hampshire; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire.
  • Gladstone DJ; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; Department of Medicine, Radiation Oncology, Geisel School of Medicine, Dartmouth College Hanover, New Hampshire; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire.
  • Williams BB; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; Department of Medicine, Radiation Oncology, Geisel School of Medicine, Dartmouth College Hanover, New Hampshire; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire.
  • Chen E; Cheshire Medical Center, Keene, New Hampshire.
  • Dexter CA; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire.
  • Thompson L; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire.
  • Bruza P; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire.
  • Pogue BW; Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire; Department of Surgery, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire; Department of Medical Physics, Wisconsin Inst
Pract Radiat Oncol ; 13(2): 153-165, 2023.
Article em En | MEDLINE | ID: mdl-36375771
PURPOSE: The use of a linear accelerator (LINAC) in ultrahigh-dose-rate (UHDR) mode can provide a conduit for wider access to UHDR FLASH effects, sparing normal tissue, but care needs to be taken in the use of such systems to ensure errors are minimized. The failure mode and effects analysis was carried out in a team that has been involved in converting a LINAC between clinical use and UHDR experimental mode for more than 1 year after the proposed methods of TG100. METHODS AND MATERIALS: A team of 9 professionals with extensive experience were polled to outline the process map and workflow for analysis, and developed fault trees for potential errors, as well as failure modes that would result. The team scored the categories of severity magnitude, occurrence likelihood, and detectability potential in a scale of 1 to 10, so that a risk priority number (RPN = severity×occurrence×detectability) could be assessed for each. RESULTS: A total of 46 potential failure modes were identified, including 5 with an RPN >100. These failure modes involved (1) patient set up, (2) gating mechanisms in delivery, and (3) detector in the beam stop mechanism. The identified methods to mitigate errors included the (1) use of a checklist post conversion, (2) use of robust radiation detectors, (3) automation of quality assurance and beam consistency checks, and (4) implementation of surface guidance during beam delivery. CONCLUSIONS: The failure mode and effects analysis process was considered critically important in this setting of a new use of a LINAC, and the expert team developed a higher level of confidence in the ability to safely move UHDR LINAC use toward expanded research access.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Análise do Modo e do Efeito de Falhas na Assistência à Saúde Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Pract Radiat Oncol Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Análise do Modo e do Efeito de Falhas na Assistência à Saúde Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Pract Radiat Oncol Ano de publicação: 2023 Tipo de documento: Article