Your browser doesn't support javascript.
loading
Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients.
Blanchet, Marie-Anne; Mercier, Gabriel; Delobel, Antoine; Nayet, Emi; Bouchard, Pierre-Alexandre; Simard, Serge; L'Her, Erwan; Branson, Richard D; Lellouche, François.
Afiliação
  • Blanchet MA; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • Mercier G; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • Delobel A; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • Nayet E; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • Bouchard PA; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • Simard S; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
  • L'Her E; Medical Intensive Care Unit, CHRU Brest, La Cavale Blanche, Brest, France.
  • Branson RD; Department of Surgery, Division of Trauma and Critical Care, University Cincinnati, Cincinnati, Ohio.
  • Lellouche F; Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada. francois.lellouche@criucpq.ulaval.ca.
Respir Care ; 68(5): 565-574, 2023 05.
Article em En | MEDLINE | ID: mdl-36596654
ABSTRACT

BACKGROUND:

An accurate SpO2 value is critical in order to optimally titrate oxygen delivery to patients and to follow oxygenation guidelines. Limited prospective data exist on real-world performance of pulse oximeters in critically ill patients. The objective of this study was to assess accuracy and bias of the SpO2 values measured by several oximeters in hospitalized subjects.

METHODS:

We included stable adults in the ICU with an arterial catheter in place. Main exclusion criteria were poor SpO2 signal and SpO2 > 96%. In each subject, we simultaneously evaluated 4 oximeters Nonin (Plymouth, Minnesota) embedded in the FreeO2 device (OxyNov, Québec City, Québec, Canada), Masimo (Radical-7, Masimo, Irvine, California), Philips (FAST, Philips, Amsterdam, the Netherlands), and Nellcor (N-600, Medtronic, Minneapolis, Minnesota). Arterial blood gases were drawn and simultaneously each oximeters' SpO2 values were collected. SpO2 values were compared to the reference (arterial oxygen saturation [SaO2 ] value) to determine bias and accuracy. The ability for oximeters to detect hypoxemia and the impact of oximeters on oxygen titration were evaluated.

RESULTS:

We included 193 subjects (153 male, mean age 66 y) in whom 211 sets of measurements were performed. The skin pigmentation evaluated by Fitzpatrick scale showed 96.2% of subjects were light skin (types 1 and 2). One oximeter overestimated SaO2 (Philips, +0.9%), whereas the 3 others underestimated SaO2 (Nonin -3.1%, Nellcor -0.3%, Masimo -0.2%). SaO2 was underestimated with Nonin oximeter in 91.3% of the cases, whereas it was overestimated in 55.2% of the cases with Philips oximeter. Moderate hypoxemia (SaO2 86-90% or PaO2 55-60 mm Hg) was detected in 92, 33, 42, and 11% of the cases with Nonin, Nellcor, Masimo, and Philips, respectively.

CONCLUSIONS:

We found significant bias and moderate accuracy between the tested oximeters and the arterial blood gases in the studied population. These discrepancies may have important clinical impact on the detection of hypoxemia and management of oxygen therapy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transtornos Respiratórios / Estado Terminal Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Humans / Male Idioma: En Revista: Respir Care Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transtornos Respiratórios / Estado Terminal Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Humans / Male Idioma: En Revista: Respir Care Ano de publicação: 2023 Tipo de documento: Article