Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets.
Comuzzi, Luca; Tumedei, Margherita; Romasco, Tea; Petrini, Morena; Afrashtehfar, Kelvin I; Inchingolo, Francesco; Piattelli, Adriano; Di Pietro, Natalia.
Afiliação
  • Comuzzi L; Independent Researcher, San Vendemiano-Conegliano Veneto, 31020 Treviso, Italy.
  • Tumedei M; Department of Medical, Surgical, and Dental Sciences, University of Milan, 20122 Milan, Italy.
  • Romasco T; Department of Medical, Oral and Biotechnological Sciences, "G. d'Annunzio" University of Chieti-Pescara, 66013 Chieti, Italy.
  • Petrini M; Center for Advanced Studies and Technology-CAST, "G. d'Annunzio" University of Chieti-Pescara, 66013 Chieti, Italy.
  • Afrashtehfar KI; Department of Medical, Oral and Biotechnological Sciences, "G. d'Annunzio" University of Chieti-Pescara, 66013 Chieti, Italy.
  • Inchingolo F; Clinical Sciences Department, College of Dentistry, Ajman University, Ajman Emirate P.O. Box 346, United Arab Emirates.
  • Piattelli A; Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, 3010 Berne, Switzerland.
  • Di Pietro N; Department of Interdisciplinary Medicine, University of Bari "Aldo Moro", 70124 Bari, Italy.
J Funct Biomater ; 14(1)2022 Dec 23.
Article em En | MEDLINE | ID: mdl-36662057
ABSTRACT
Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws.