Your browser doesn't support javascript.
loading
Application of liquid scintillation light guide (LSLG) to scattered X-ray measurement from dental panoramic radiography system.
Nomura, Kiyoshi; Takahashi, Masashi; Fujibuchi, Toshioh; Koike, Yuya; Hara, Masayuki; Higaki, Shogo; Kobayashi, Ikuo; Yoshino, Souichirou.
Afiliação
  • Nomura K; Radio-Isotope Research Laboratory, Tokyo Medical University, Shinjuku 6-1-1, Shinjuku-ku, Tokyo, 160-8402, Japan. Electronic address: k-nomura@tokyo-med.ac.jp.
  • Takahashi M; Radio-Isotope Research Laboratory, Tokyo Medical University, Shinjuku 6-1-1, Shinjuku-ku, Tokyo, 160-8402, Japan.
  • Fujibuchi T; Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Maidashi 3-1-1, Higashi-ku, Fukuoka, 812-8582, Japan.
  • Koike Y; Faculty of Science and Engineering, Meiji University, Higashi-mita 1-1-1, Tama-ku, Kawasaki, Kanagawa, 214-8571, Japan.
  • Hara M; Radioisotope Center, Tokyo Medical and Dental University (TMDU), Yushima 1-5-45, Bunkyo-ku, Tokyo, 113-8510, Japan.
  • Higaki S; Isotope Science Center, The University of Tokyo, Yayoi 2-11-16, Bunkyou-ku, Tokyo, 113-0032, Japan.
  • Kobayashi I; Research Institute of Nuclear Engineering, University of Fukui, Kanawa-cho1-3-33, Tsuruga, Fukui, 914-0055, Japan.
  • Yoshino S; Yoshino Dental Clinic, Yayoidai 25-1, Izumi-ku, Yokohama, Kanagawa, 245-0008, Japan.
Appl Radiat Isot ; 194: 110720, 2023 Apr.
Article em En | MEDLINE | ID: mdl-36787680
ABSTRACT
A 1.2 m flexible liquid scintillation light guide (LSLG) detector connected to a portable electric device was developed and applied to monitor X-rays scattered from a panoramic dental X-ray imaging apparatus. The X-ray absorption of the LSLG tube was simulated for 20 and 40 keV X-rays. The LSLG detector was calibrated by comparing doses measured by a 3-inch NaI(Tl) detector using Am-241 and Cs-137 sources. A linear relationship was obtained between dose rates (µSv/h) and count rates (cps). The elapsed time profile of scattered radiation from a panoramic dental X-ray imaging apparatus was determined. Local absorbed doses were measured with small OSL dosimeters on the LSLG tube suspended over the shoulder and the eyeglass. Behind the ears and neck (in the center of the LSLG tube) showed high doses. The LSLG detector was useful for real-time monitoring of scattered X-rays.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioisótopos de Césio Tipo de estudo: Diagnostic_studies Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioisótopos de Césio Tipo de estudo: Diagnostic_studies Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2023 Tipo de documento: Article