Clinical assessment of SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay.
Pighi, Laura; Henry, Brandon M; De Nitto, Simone; Gianfilippi, Gianluca; Salvagno, Gian Luca; Plebani, Mario; Lippi, Giuseppe.
Afiliação
  • Pighi L; Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.
  • Henry BM; Service of Laboratory Medicine, Pederzoli Hospital, Peschiera del Garda, Italy.
  • De Nitto S; Clinical Laboratory, Division of Nephrology and Hypertension, Cincinnati Children's Hospital Medical Center, Cincinnati OH, USA.
  • Gianfilippi G; Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.
  • Salvagno GL; Service of Laboratory Medicine, Pederzoli Hospital, Peschiera del Garda, Italy.
  • Plebani M; Medical Direction, Pederzoli Hospital, Peschiera del Garda, Italy.
  • Lippi G; Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.
Clin Chem Lab Med ; 61(8): 1506-1510, 2023 07 26.
Article em En | MEDLINE | ID: mdl-36883744
ABSTRACT

OBJECTIVES:

Given that SARS-CoV-2 antigen tests will represent a pillar for supporting or surrogating molecular testing in the endemic period, we report here the clinical performance of the new SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay (MAG-CLIA SARS-CoV-2 Ag).

METHODS:

The study population consisted of 181 subjects (mean age 61 ± 21 years; 92 females) undergoing coronavirus disease 2019 (COVID-19) testing at the local diagnostic facility, from December 2022 to February 2023. Routine diagnostic practice involved the collection of a double nostril nasopharyngeal swab, analyzed in duplicate with SARS-CoV-2 antigen (MAG-CLIA SARS-CoV-2 Ag) and molecular (Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit) tests.

RESULTS:

A significant Spearman's correlation was found between MAG-CLIA SARS-CoV-2 Ag and mean Ct values of SARS-CoV-2 E and S genes (r=-0.95; p<0.001). In all nasopharyngeal samples, the area under the curve (AUC) of MAG-CLIA SARS-CoV-2 Ag was 0.86 (95% CI, 0.81-0.90), with 0.71 sensitivity and 1.00 specificity at 7 ng/L cut-off, increasing to 0.98 (95% CI, 0.96-1.00) AUC and 0.96 sensitivity (with 0.97 specificity) in high viral load samples. When SARS-CoV-2 N protein concentration was replaced with raw instrumental readings (i.e., relative light units [RLU]), the AUC in all samples increased to 0.94. A RLU value of 945 was associated with 88.4% accuracy, 0.85 sensitivity, 0.95 specificity, 0.77 negative predictive value (NPV) and 0.97 positive predictive value (PPV), respectively.

CONCLUSIONS:

We found satisfactory analytical performance of MAG-CLIA SARS-CoV-2 Ag, which could be used as surrogate of molecular testing for identifying high viral load samples. Broadening the reportable range of values may generate even better performance.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: SARS-CoV-2 / COVID-19 Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: SARS-CoV-2 / COVID-19 Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2023 Tipo de documento: Article