In vivo quantitative MRI: T1 and T2 measurements of the human brain at 0.064 T.
Jordanova, Kalina V; Martin, Michele N; Ogier, Stephen E; Poorman, Megan E; Keenan, Kathryn E.
Afiliação
  • Jordanova KV; Physical Measurement Laboratory, National Institute of Standards and Technology, NIST, Boulder, CO, USA. kalina.jordanova@nist.gov.
  • Martin MN; Physical Measurement Laboratory, National Institute of Standards and Technology, NIST, Boulder, CO, USA.
  • Ogier SE; Physical Measurement Laboratory, National Institute of Standards and Technology, NIST, Boulder, CO, USA.
  • Poorman ME; Department of Physics, University of Colorado Boulder, Boulder, CO, USA.
  • Keenan KE; Hyperfine, Inc., Guilford, CT, USA.
MAGMA ; 36(3): 487-498, 2023 Jul.
Article em En | MEDLINE | ID: mdl-37208553
OBJECTIVE: To measure healthy brain [Formula: see text] and [Formula: see text] relaxation times at 0.064 T. MATERIALS AND METHODS: [Formula: see text] and [Formula: see text] relaxation times were measured in vivo for 10 healthy volunteers using a 0.064 T magnetic resonance imaging (MRI) system and for 10 test samples on both the MRI and a separate 0.064 T nuclear magnetic resonance (NMR) system. In vivo [Formula: see text] and [Formula: see text] values are reported for white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) for automatic segmentation regions and manual regions of interest (ROIs). RESULTS: [Formula: see text] sample measurements on the MRI system were within 10% of the NMR measurement for 9 samples, and one sample was within 11%. Eight [Formula: see text] sample MRI measurements were within 25% of the NMR measurement, and the two longest [Formula: see text] samples had more than 25% variation. Automatic segmentations generally resulted in larger [Formula: see text] and [Formula: see text] estimates than manual ROIs. DISCUSSION: [Formula: see text] and [Formula: see text] times for brain tissue were measured at 0.064 T. Test samples demonstrated accuracy in WM and GM ranges of values but underestimated long [Formula: see text] in the CSF range. This work contributes to measuring quantitative MRI properties of the human body at a range of field strengths.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Substância Branca Limite: Humans Idioma: En Revista: MAGMA Ano de publicação: 2023 Tipo de documento: Article