Your browser doesn't support javascript.
loading
Introduction of an electrochemical point-of-care assay for quantitative determination of paracetamol in finger-prick capillary whole blood samples.
Kujala, Johanna; Wester, Niklas; Lohela, Terhi J; Kurkela, Mika; Backman, Janne T; Mikladal, Björn; Laurila, Tomi; Koskinen, Jari; Lilius, Tuomas O; Kalso, Eija A.
Afiliação
  • Kujala J; Department of Emergency Medicine and Services, Helsinki University Hospital and University of Helsinki, Helsinki, Finland.
  • Wester N; Department of Chemistry and Materials Science, School of Chemical Engineering, Aalto University, Helsinki, Finland.
  • Lohela TJ; Department of Electrical Engineering and Automation, School of Electrical Engineering, Aalto University, Helsinki, Finland.
  • Kurkela M; Fepod Oy Ltd, c/o Terkko Health Hub, Helsinki, Finland.
  • Backman JT; Department of Clinical Pharmacology, Faculty of Medicine, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
  • Mikladal B; Department of Anaesthesiology, Intensive Care and Pain Medicine, Helsinki University Hospital and University of Helsinki, Helsinki, Finland.
  • Laurila T; Individualized Drug Therapy Research Program, Faculty of Medicine, University of Helsinki, Helsinki, Finland.
  • Koskinen J; Department of Clinical Pharmacology, Faculty of Medicine, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
  • Lilius TO; Individualized Drug Therapy Research Program, Faculty of Medicine, University of Helsinki, Helsinki, Finland.
  • Kalso EA; Department of Clinical Pharmacology, Faculty of Medicine, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Br J Clin Pharmacol ; 89(9): 2933-2938, 2023 09.
Article em En | MEDLINE | ID: mdl-37218304
ABSTRACT

AIMS:

Measuring venous plasma paracetamol concentrations is time- and resource-consuming. We aimed to validate a novel electrochemical point-of-care (POC) assay for rapid paracetamol concentration determinations.

METHODS:

Twelve healthy volunteers received 1 g oral paracetamol, and its concentrations were analysed 10 times over 12 h for capillary whole blood (POC), venous plasma (high-performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS)), and dried capillary blood (HPLC-MS/MS).

RESULTS:

At concentrations >30 µM, POC showed upward biases of 20% (95% limits of agreement [LOA] -22 to 62) and 7% (95% LOA -23 to 38) compared with venous plasma and capillary blood HPLC-MS/MS, respectively. There were no significant differences between mean concentrations for the paracetamol elimination phase.

CONCLUSIONS:

Upward biases in POC compared with venous plasma HPLC-MS/MS were likely due to higher paracetamol concentrations in capillary blood than in venous plasma and to faulty individual sensors. The novel POC method is a promising tool for paracetamol concentration analysis.