A Planar Disk Electrode Chip Based on MWCNT/CS/Pb2+ Ionophore IV Nanomaterial Membrane for Trace Level Pb2+ Detection.
Molecules
; 28(10)2023 May 17.
Article
em En
| MEDLINE
| ID: mdl-37241883
Unlike conventional lead ion (Pb2+) detecting methods, electrochemical methods have the attractive advantages of rapid response, good portability and high sensitivity. In this paper, a planar disk electrode modified by multiwalled carbon nanotube (MWCNTs)/chitosan (CS)/lead (Pb2+) ionophore IV nanomaterial and its matched system are proposed. This system presented a good linear relationship between the concentration of Pb2+ ions and the peak current in differential pulse stripping voltammetry (DPSV), under optimized conditions of -0.8 V deposition potential, 5.5 pH value, 240 s deposition time, performed sensitive detection of Pb2+ within sensitivity of 1.811 µA · µg-1 and detection limit of 0.08 µg · L-1. Meanwhile, the results of the system in detecting lead ions in real seawater samples are highly similar to that of inductively coupled plasma emission spectrometer (ICP-MS), which proved a practicability for the system in detection of trace-level Pb2+.