Impact of list-mode reconstruction and image-space point spread function correction on PET image contrast and quantitative value using SiPM-based PET/CT system.
Shirakawa, Yuya; Matsutomo, Norikazu.
Afiliação
  • Shirakawa Y; Department of Radiology, Kyorin University Hospital, 6-20-2 Shinkawa, Mitaka, Tokyo, 181-8611, Japan. yuya_shirakawa@ks.kyorin-u.ac.jp.
  • Matsutomo N; Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University, Mitaka, Tokyo, 181-8612, Japan.
Radiol Phys Technol ; 16(3): 384-396, 2023 Sep.
Article em En | MEDLINE | ID: mdl-37368168
ABSTRACT
We evaluate the effects of list-mode reconstruction and the image-space point spread function (iPSF) on the contrast and quantitative values of positron emission tomography (PET) images using a SiPM-PET/CT system. The evaluation is conducted on an NEMA body phantom and clinical images using a Cartesion Prime SiPM-PET/CT system. The signal-to-background ratio (SBR) of the phantom is set to 2, 4, 6, and 8, and all the PET image data are obtained and reconstructed using 3D-OSEM, time-of-flight, iPSF (-/ +), and a 4-mm Gaussian filter with several iterations. The evaluation criteria include % background variability (NB,10 mm), % contrast (QH,10 mm), iPSF change in QH,10 mm (ΔQH,10 mm) for edge artifact evaluation, profile curves, visual evaluation of edge artifacts, clinical imaging for the standardized uptake value (SUV) of lung nodules, and SNRliver. NB,10 mm demonstrates no significant difference in all SBRs with and without iPSF, whereas QH,10 mm is higher based on the SBR with and without iPSF. ΔQH,10 mm indicates increased iterations and a larger rate of change (> 5%) for small spheres of < 17 mm. The profile curves portrayed almost real concentrations, except for the 10-mm sphere of SBR2 without iPSF; however, with iPSF, an overshoot was observed in the 13-mm sphere of all SBRs. The degree of overshoot increased with increasing iteration and SBR. Edge artifacts were detected at values ≥ 17-22 mm in SBRs other than SBR2 with iPSF. Irrespective of the nodal size, SUV and SNRliver improved considerably after iPSF adjustment. Therefore, the effects of list-mode reconstruction and iPSF on PET image contrast were limited, and the overcorrection of the quantitative values was validated using iPSF.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada Idioma: En Revista: Radiol Phys Technol Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada Idioma: En Revista: Radiol Phys Technol Ano de publicação: 2023 Tipo de documento: Article