Validation of the oscillometric home blood pressure monitor Braun BUA4000 with wide-range cuff in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.
Ntousopoulos, Vasileios; Ntineri, Angeliki; Theodosiadi, Aikaterini; Stathopoulou, Panagiota; Menti, Ariadni; Kyriakoulis, Konstantinos G; Kollias, Anastasios; Stergiou, George S.
Afiliação
  • Ntousopoulos V; Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece.
Blood Press Monit ; 29(2): 99-102, 2024 Apr 01.
Article em En | MEDLINE | ID: mdl-37997431
OBJECTIVE: To evaluate the blood pressure (BP) measurement accuracy of the Braun BUA4000 automated oscillometric upper-arm cuff device for self-home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01. METHODS: Participants were recruited to fulfill the age, sex, BP and limb circumference distribution criteria of the AAMI/ESH/ISO Universal Standard in a general population using the same arm sequential BP measurement method. A single wide-range cuff of the test device was used for arm circumference range 22-42 cm. RESULTS: A total of 98 individuals were recruited and 85 were analyzed [mean age 60.3 ±â€…16.1 (SD) years, 44 men, arm circumference 31.5 ±â€…5.1 cm, range 22-41.5 cm]. For validation criterion 1, the mean difference ± SD between the test device and reference BP readings ( N  = 255) was 0.9 ±â€…6.4/-0.3 ±â€…6.4 mmHg (systolic/diastolic; threshold ≤5 ±â€…8 mmHg). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual ( N  = 85) was 5.15/5.81 mmHg (systolic/diastolic; threshold ≤6.88/6.95 mmHg). CONCLUSION: The Braun BUA4000 automated oscillometric BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard in a general population and can be recommended for self-monitoring of BP by patients at home.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitores de Pressão Arterial / Hipertensão Limite: Adult / Aged / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitores de Pressão Arterial / Hipertensão Limite: Adult / Aged / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2024 Tipo de documento: Article