Measurement of energy and directional distribution of neutron ambient dose equivalent for the 7Li(p,n)7Be reaction.
Appl Radiat Isot
; 204: 111140, 2024 Feb.
Article
em En
| MEDLINE
| ID: mdl-38070360
ABSTRACT
Double differential neutron fluence distributions were measured in the 7Li(p,n)7Be reaction for proton beam energies 7, 9 and 12 MeV. Seven liquid scintillator based detectors were employed to measure neutron fluence distributions using the Time of Flight technique. Neutron ambient dose equivalents were determined from the measured fluence distribution using ICRP (International Commission on Radiological Protection) recommended fluence to dose equivalent conversion coefficients. Neutron dose equivalents were also measured using a conventional BF3 detector based REM counter. Ambient dose equivalent measured by the REM counter is found to be in agreement with that determined from the neutron fluence spectra within their uncertainties. Angular distributions of the ambient dose equivalents were also determined from the measured fluence distributions at different angles.