Your browser doesn't support javascript.
loading
Automatic marker-based alignment method for a nano-resolution full-field transmission X-ray microscope.
Appl Opt ; 62(36): 9536-9543, 2023 Dec 20.
Article em En | MEDLINE | ID: mdl-38108778
ABSTRACT
Driven by the development of X-ray optics, the spatial resolution of the full-field transmission X-ray microscope (TXM) has reached tens of nanometers and plays an important role in promoting the development of biomedicine and materials science. However, due to the thermal drift and the radial/axial motion error of the rotation stage, TXM computed tomography (CT) data are often associated with random image jitter errors along the horizontal and vertical directions during CT measurement. A nano-resolution 3D structure information reconstruction is almost impossible without a prior appropriate alignment process. To solve this problem, a fully automatic gold particle marker-based alignment approach without human intervention was proposed in this study. It can automatically detect, isolate, and register gold particles for projection image alignment with high efficiency and accuracy, facilitating a high-quality tomographic reconstruction. Simulated and experimental results confirmed the reliability and robustness of this method.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Opt Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Opt Ano de publicação: 2023 Tipo de documento: Article