The influence of truncated-conical implant length on primary stability in maxillary and mandibular regions: an in vitro study using polyurethane blocks.
Romasco, Tea; Pignatelli, Pamela; Tumedei, Margherita; Hossein, Hamid Heydari Sheikh; Cipollina, Alessandro; Piattelli, Adriano; Inchingolo, Francesco; Di Pietro, Natalia.
Afiliação
  • Romasco T; Center for Advanced Studies and Technology-CAST, "G. D'Annunzio, University of Chieti-Pescara, Via Luigi Polacchi 11, 66100, Chieti, Italy.
  • Pignatelli P; Department of Medical, Oral and Biotechnological Sciences, "G. D'Annunzio" University of Chieti-Pescara, Via dei Vestini 31, 66100, Chieti, Italy.
  • Tumedei M; Department of Medical, Oral and Biotechnological Sciences, "G. D'Annunzio" University of Chieti-Pescara, Via dei Vestini 31, 66100, Chieti, Italy.
  • Hossein HHS; Department of Medical, Surgical, and Dental Sciences, University of Milan, Via della Commenda 10, 20122, Milan, Italy.
  • Cipollina A; Center for Advanced Studies and Technology-CAST, "G. D'Annunzio, University of Chieti-Pescara, Via Luigi Polacchi 11, 66100, Chieti, Italy.
  • Piattelli A; Department of Medical, Oral and Biotechnological Sciences, "G. D'Annunzio" University of Chieti-Pescara, Via dei Vestini 31, 66100, Chieti, Italy.
  • Inchingolo F; Villa Serena Foundation for Research, Via Leonardo Petruzzi 42, 65013, Città Sant'Angelo, Italy.
  • Di Pietro N; , Via Piacenza 7, 92019, Sciacca, Italy.
Clin Oral Investig ; 28(1): 28, 2023 Dec 26.
Article em En | MEDLINE | ID: mdl-38147179
ABSTRACT

OBJECTIVES:

This in vitro study is aimed at assessing whether implant primary stability is influenced by implant length in artificial bone with varying densities. MATERIALS AND

METHODS:

A total of 120 truncated-conical implants (60 long-length 3p L, 3.8 × 14 mm; 60 short-length 3p S, 3.8 × 8 mm) were inserted into 20, 30, and 40 pounds per cubic foot (PCF) density polyurethane blocks. The insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values were recorded for each experimental condition.

RESULTS:

In 30 and 40 PCF blocks, 3p S implants exhibited significantly higher IT values (90 and 80 Ncm, respectively) than 3p L (85 and 50 Ncm, respectively). Similarly, RT was significantly higher for 3p S implants in 30 and 40 PCF blocks (57 and 90 Ncm, respectively). However, there were no significant differences in RFA values, except for the 20 PCF block, where 3pS implants showed significantly lower values (63 ISQ) than 3p L implants (67 ISQ) in both the distal and mesial directions.

CONCLUSIONS:

These results demonstrated that the implant's length mainly influences the IT and RT values in the polyurethane blocks that mimic the mandibular region of the bone, resulting in higher values for the 3p S implants, while the RFA values remained unaffected. However, in the lowest density block simulating the maxillary bone, 3p L implants exhibited significantly higher ISQ values. CLINICAL RELEVANCE Therefore, our data offer valuable insights into the biomechanical behavior of these implants, which could be clinically beneficial for enhancing surgical planning.