Does radiographic calibration affect linear radiographic measurements in a large pediatric spine registry?
Regan, Christina M; Morgan, Jessica; Hilaire, Tricia St; El-Hawary, Ron; Pahys, Joshua M; Skaggs, David L; Yaszay, Burt; Parent, Stefan; Larson, A Noelle.
Afiliação
  • Regan CM; Department of Orthopedic Surgery, Mayo Clinic, 200 First Street SW, Rochester, MN, USA.
  • Morgan J; Pediatric Spine Study Group, Valley Forge, PA, USA.
  • Hilaire TS; Pediatric Spine Study Group, Valley Forge, PA, USA.
  • El-Hawary R; Dalhousie University, Halifax, NS, Canada.
  • Pahys JM; Philadelphia Shriners Hospital for Children, Philadelphia, PA, USA.
  • Skaggs DL; Cedars-Sinai Health System, Los Angeles, CA, USA.
  • Yaszay B; Seattle Children's Hospital, Seattle, WA, USA.
  • Parent S; St. Justine's Hospital, Montreal, QC, Canada.
  • Larson AN; Department of Orthopedic Surgery, Mayo Clinic, 200 First Street SW, Rochester, MN, USA. larson.noelle@mayo.edu.
Spine Deform ; 12(2): 335-339, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38147269
ABSTRACT

PURPOSE:

Large registries are increasingly at the forefront of modern pediatric spine research, with manual, centralized, trained radiographic measurement serving as the gold standard for spine research. However, there is limited data regarding the reliability of registry measurements which may be subject to differences in radiographic calibration. We undertook this study to evaluate reliability of T1-T12 height, L1-S1 height, and coronal balance measurements for a large registry of early onset scoliosis patients.

METHODS:

Three trained technicians from the Pediatric Spine Study Group measured 43 radiographs for T1-T12, L1-S1, and coronal balance using 3 different calibration techniques. All radiographs were AP views of patients with magnetically controlled growing rods with known diameters. The calibration techniques used a pre-export manually drawn line, a digital automatically generated computerized marker, and the diameter of a magnetically controlled growing rod. The intraclass correlation coefficient (ICC) was calculated to determine reliability.

RESULTS:

1161 measurements were performed. For each of the three raters, coronal balance, T1-T12 height and L1-S1 height had excellent agreement regardless of the calibration technique (alpha 0.93-1.0). Among the parameters, coronal balance had the worst inter-rater reliability, whereas there was excellent interrater reliability regarding T1-T12 height and L1-S1 height (alpha 0.91-0.99).

CONCLUSION:

There was excellent agreement among reviewers and between the 3 different calibration techniques. While calibration using rod diameter served as the gold standard, this data shows that other standard calibration methods were adequate and achieved excellent reliability for registry radiographs.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Escoliose / Coluna Vertebral Limite: Child / Humans Idioma: En Revista: Spine Deform Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Escoliose / Coluna Vertebral Limite: Child / Humans Idioma: En Revista: Spine Deform Ano de publicação: 2024 Tipo de documento: Article