Analytical evaluation of the novel Mindray high sensitivity cardiac troponin I immunoassay on CL-1200i.
Lippi, Giuseppe; Pighi, Laura; Paviati, Elisa; Demonte, Davide; De Nitto, Simone; Gelati, Matteo; Montagnana, Martina; Gandini, Giorgio; Henry, Brandon M; Salvagno, Gian Luca.
Afiliação
  • Lippi G; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Pighi L; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Paviati E; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Demonte D; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • De Nitto S; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Gelati M; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Montagnana M; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
  • Gandini G; Transfusion Center, University Hospital of Verona, Verona, Italy.
  • Henry BM; Clinical Laboratory, Division of Nephrology and Hypertension, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Salvagno GL; Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
Clin Chem Lab Med ; 62(7): 1433-1437, 2024 Jun 25.
Article em En | MEDLINE | ID: mdl-38176023
ABSTRACT

OBJECTIVES:

The current study was designed to evaluate the analytical performance of the new Mindray highly sensitive cardiac troponin I (hs-cTnI) chemiluminescent immunoassay on Mindray CL-1200i, as a thorough validation of novel hs-cTnI methods is required before introduction into clinical practice.

METHODS:

The evaluation of the analytical performance of this hs-cTnI immunoassay encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity, 99th percentile upper reference limit (URL) and concordance with another previously validated hs-cTnI chemiluminescent immunoassay.

RESULTS:

The LOB and LOD were 0.32 and 0.35 ng/L, whilst the functional sensitivity (expressed as cTnI value with <10 % imprecision), was 0.35 ng/L. The linearity was excellent throughout a wide range of clinically measurable values (r=1.00 between 0.8 and 9,726.9 ng/mL). The intra-assay, inter-assay and total imprecision were 1.1-1.3 %, 5.5-8.1 % and 5.6-8.2 %, respectively. The 99th percentile URL calculated using residual plasma from 246 ostensibly healthy blood donors was 9.2 ng/L (4.3 ng/L in women vs. 12.3 ng/L in men). The Spearman's correlation between Mindray hs-cTnI and Access hs-TnI was 0.97, with mean bias of 7.2 % (95 % CI, 2.6-11.9 %).

CONCLUSIONS:

Although we failed to confirm the very optimistic analytical characteristics previously reported for this method, our evaluation of the novel Mindray hs-cTnI immunoassay on CL-1200i demonstrated that the overall performance is comparable to that of other commercially available hs-cTnI techniques, making it a viable alternative to other methods.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Troponina I / Limite de Detecção Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Troponina I / Limite de Detecção Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2024 Tipo de documento: Article