Feasibility and Effectiveness of a Ring-Type Blood Pressure Measurement Device Compared With 24-Hour Ambulatory Blood Pressure Monitoring Device.
Lee, Huijin; Park, Sungjoon; Kwon, Hyuktae; Cho, Belong; Park, Jin Ho; Lee, Hae-Young.
Afiliação
  • Lee H; Division of Cardiology, Department of Internal Medicine, Seoul National University Hospital, Seoul, Korea.
  • Park S; Division of Cardiology, Department of Internal Medicine, Seoul National University Hospital, Seoul, Korea.
  • Kwon H; Department of Family Medicine, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Korea.
  • Cho B; Department of Family Medicine, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Korea.
  • Park JH; Department of Family Medicine, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Korea.
  • Lee HY; Division of Cardiology, Department of Internal Medicine, Seoul National University Hospital, Seoul, Korea.
Korean Circ J ; 54(2): 93-104, 2024 Feb.
Article em En | MEDLINE | ID: mdl-38196118
ABSTRACT
BACKGROUNDS AND

OBJECTIVES:

This study aimed to evaluate the applicability and precision of a ring-type cuffless blood pressure (BP) measurement device, CART-I Plus, compared to conventional 24-hour ambulatory BP monitoring (ABPM).

METHODS:

Forty patients were recruited, and 33 participants were included in the final analysis. Each participant wore both CART-I Plus and ABPM devices on the same arm for approximately 24 hours. BP estimation from CART-I Plus, derived from photoplethysmography (PPG) signals, were compared with the corresponding ABPM measurements.

RESULTS:

The CART-I Plus recorded systolic blood pressure (SBP)/diastolic blood pressure (DBP) values of 131.4±14.1/81.1±12.0, 132.7±13.9/81.9±11.9, and 128.7±14.6/79.3±12.2 mmHg for 24-hour, daytime, and nighttime periods respectively, compared to ABPM values of 129.7±11.7/84.4±11.2, 131.9±11.6/86.3±11.1, and 124.5±13.6/80.0±12.2 mmHg. Mean differences in SBP/DBP between the two devices were 1.74±6.69/-3.24±6.51 mmHg, 0.75±7.44/-4.41±7.42 mmHg, and 4.15±6.15/-0.67±5.23 mmHg for 24-hour, daytime, and nighttime periods respectively. Strong correlations were also observed between the devices, with r=0.725 and r=0.750 for transitions in SBP and DBP from daytime to nighttime, respectively (both p<0.001).

CONCLUSIONS:

The CART-I Plus device, with its unique ring-type design, shows promising accuracy in BP estimation and offers a potential avenue for continuous BP monitoring in clinical practice. TRIAL REGISTRATION ClinicalTrials.gov Identifier NCT06084065.