The influence of flowmeters on rhinomanometry results and detection of nasal airflow asymmetry.
Otolaryngol Pol
; 78(1): 1-7, 2024 Feb 29.
Article
em En
| MEDLINE
| ID: mdl-38332709
ABSTRACT
<b><br>Introduction:
</b> Rhinomanometry is an otolaryngological diagnostic method used to determine airflow as a function of the pressure drop through the left and right nasal cavities. Airflow is measured using orifice flowmeters that attenuate the flow.</br> <b><br>Aim:
</b> This paper describes the results of a study into the effects of flowmeter design on rhinomanometry results and detection of nasal airflow asymmetry.</br> <b><br>Material andmethods:
</b> Four flowmeters were examined using a 3D printed model of a human nose.</br> <b><br>Conclusions:
</b> Each flowmeter interfered with the rhinomanometry results.</br>.Palavras-chave