High-contrast handheld FFOCT system for tissue imaging.
Opt Lett ; 49(4): 1081-1084, 2024 Feb 15.
Article em En | MEDLINE | ID: mdl-38359258
ABSTRACT
A low contrast is a limiting factor for imaging a microstructure beneath the biological sample surface. In this work, we describe a novel, to our knowledge, full-field optical coherence tomography (FFOCT) system with a probe connected by a fiber bundle and a multimode optical fiber. The device is based on the tandem structure of the Michelson interferometer and the Fizeau interferometer. One advantage of our device is that light propagates through the fiber bundle only once, greatly improving detection sensitivity. In addition, by spatial filtering in the Fourier domain and inverse filtering, the effects of pixelation artifacts and multiple scattering in the en face images obtained by our system are suppressed. The depth-resolved en face images of the human finger skin ex vivo and the porcine esophagus ex vivo are presented to demonstrate the capability of our system.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pele / Tomografia de Coerência Óptica Limite: Animals / Humans Idioma: En Revista: Opt Lett Ano de publicação: 2024 Tipo de documento: Article