Monitoring of Remotely Reprogrammable Implantable Loop Recorders With Algorithms to Reduce False-Positive Alerts.
Neiman, Zachary M; Raitt, Merritt H; Rohrbach, Gregory; Dhruva, Sanket S.
Afiliação
  • Neiman ZM; University of California, San Francisco School of Medicine San Francisco CA USA.
  • Raitt MH; Portland Veterans Affairs Health Care System Knight Cardiovascular Institute, Oregon Health and Sciences University Portland OR USA.
  • Rohrbach G; San Francisco Veterans Affairs Medical Center San Francisco CA USA.
  • Dhruva SS; University of California, San Francisco School of Medicine San Francisco CA USA.
J Am Heart Assoc ; 13(5): e032890, 2024 Mar 05.
Article em En | MEDLINE | ID: mdl-38390808
ABSTRACT

BACKGROUND:

Implantable loop recorders (ILRs) are increasingly placed for arrhythmia detection. However, historically, ≈75% of ILR alerts are false positives, requiring significant time and effort for adjudication. The LINQII and LUX-Dx are remotely reprogrammable ILRs with dual-stage algorithms using artificial intelligence to reduce false positives, but their utility in routine clinical practice has not been studied. METHODS AND

RESULTS:

We identified patients with the LINQII and LUX-Dx who were monitored by the Veterans Affairs National Cardiac Device Surveillance Program between March and June 2022. ILR programming was customized on the basis of implant indication. All alerts and every 90-day scheduled transmissions were manually reviewed. ILRs were remotely reprogrammed, as appropriate, after false-positive alerts or 2 consecutive same-type alerts, unless there was ongoing clinical need for that alert. Outcomes were total number of transmissions and false positives. We performed medical record review to determine if patients experienced any adverse clinical events, including hospitalization and mortality. Among 117 LINQII patients, there were 239 total alerts, 43 (18.0%) of which were false positives. Among 105 LUX-Dx patients, there were 300 total alerts, 115 (38.3%) of which were false positives. LINQIIs were reprogrammed 22 times, resulting in a decrease in median alerts/day from 0.13 to 0.03. LUX-Dx ILRs were reprogrammed 52 times, resulting in a decrease from 0.15 to 0.01 median alerts/day. There were no adverse clinical events that could have been identified by superior or earlier arrhythmia detection.

CONCLUSIONS:

ILRs with artificial intelligence algorithms and remote reprogramming ability are associated with reduced alert burden because of higher true-positive rates than prior ILRs, without missing potentially consequential arrhythmias.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 / 6_ODS3_enfermedades_notrasmisibles Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Inteligência Artificial Limite: Humans Idioma: En Revista: J Am Heart Assoc Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 / 6_ODS3_enfermedades_notrasmisibles Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Inteligência Artificial Limite: Humans Idioma: En Revista: J Am Heart Assoc Ano de publicação: 2024 Tipo de documento: Article