Your browser doesn't support javascript.
loading
Design, Characterization, and Preliminary Assessment of a Two-Degree-of-Freedom Powered Ankle-Foot Prosthesis.
Hsieh, Tsung-Han; Song, Hyungeun; Shu, Tony; Qiao, Junqing; Yeon, Seong Ho; Carney, Matthew; Mooney, Luke; Duval, Jean-François; Herr, Hugh.
Afiliação
  • Hsieh TH; K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Song H; Department of Media Arts and Sciences, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Shu T; K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Qiao J; Department of Media Arts and Sciences, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Yeon SH; K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Carney M; Department of Media Arts and Sciences, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Mooney L; K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Duval JF; Department of Media Arts and Sciences, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
  • Herr H; K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, MA 02142, USA.
Biomimetics (Basel) ; 9(2)2024 Jan 26.
Article em En | MEDLINE | ID: mdl-38392122
ABSTRACT
Powered ankle prostheses have been proven to improve the walking economy of people with transtibial amputation. All commercial powered ankle prostheses that are currently available can only perform one-degree-of-freedom motion in a limited range. However, studies have shown that the frontal plane motion during ambulation is associated with balancing. In addition, as more advanced neural interfaces have become available for people with amputation, it is possible to fully recover ankle function by combining neural signals and a robotic ankle. Accordingly, there is a need for a powered ankle prosthesis that can have active control on not only plantarflexion and dorsiflexion but also eversion and inversion. We designed, built, and evaluated a two-degree-of-freedom (2-DoF) powered ankle-foot prosthesis that is untethered and can support level-ground walking. Benchtop tests were conducted to characterize the dynamics of the system. Walking trials were performed with a 77 kg subject that has unilateral transtibial amputation to evaluate system performance under realistic conditions. Benchtop tests demonstrated a step response rise time of less than 50 milliseconds for a torque of 40 N·m on each actuator. The closed-loop torque bandwidth of the actuator is 9.74 Hz. Walking trials demonstrated torque tracking errors (root mean square) of less than 7 N·m. These results suggested that the device can perform adequate torque control and support level-ground walking. This prosthesis can serve as a platform for studying biomechanics related to balance and has the possibility of further recovering the biological function of the ankle-subtalar-foot complex beyond the existing powered ankles.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Biomimetics (Basel) Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Biomimetics (Basel) Ano de publicação: 2024 Tipo de documento: Article