BACKGROUND:
The fifth-generation SAPIEN 3 Ultra Resilia valve (S3UR) incorporates several design changes as compared with its predecessors, the SAPIEN 3 (S3) and SAPIEN 3 Ultra (S3U) valves, including bovine leaflets treated with a novel process intended to reduce structural valve deterioration via calcification, as well as a taller external skirt on the 29-mm valve size to reduce paravalvular leak (PVL). The clinical performance of S3UR compared with S3 and S3U in a large patient population has not been previously reported.OBJECTIVES:
The aim of this study was to compare S3UR to S3/S3U for procedural, in-hospital, and 30-day clinical and echocardiographic outcomes after transcatheter aortic valve replacement (TAVR).METHODS:
Patients enrolled in the STS/ACC TVT (Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapy) Registry between January 1, 2021, and June 30, 2023, who underwent TAVR with S3UR or S3U/S3 valve platforms were propensity-matched and evaluated for procedural, in-hospital, and 30-day clinical and echocardiographic outcomes.RESULTS:
10,314 S3UR patients were propensity matched with 10,314 patients among 150,539 S3U/S3 patients. At 30 days, there were no statistically significant differences in death, stroke, or bleeding, but a numerically higher hospital readmission rate in the S3UR cohort (8.5% vs 7.7%; P = 0.04). At discharge, S3UR patients exhibited significantly lower mean gradients (9.2 ± 4.6 mm Hg vs 12.0 ± 5.7 mm Hg; P < 0.0001) and larger aortic valve area (2.1 ± 0.7 cm2 vs 1.9 ± 0.6 cm2; P < 0.0001) than patients treated with S3/S3U. The 29-mm valve size exhibited significant reduction in mild PVL (5.3% vs 9.4%; P < 0.0001).CONCLUSIONS:
S3UR TAVR is associated with lower mean gradients and lower rates of PVL than earlier generations of balloon expandable transcatheter heart valve platforms.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Contexto em Saúde:
6_ODS3_enfermedades_notrasmisibles
Base de dados:
MEDLINE
Assunto principal:
Valva Aórtica
/
Estenose da Valva Aórtica
/
Desenho de Prótese
/
Próteses Valvulares Cardíacas
/
Sistema de Registros
/
Recuperação de Função Fisiológica
/
Valvuloplastia com Balão
/
Substituição da Valva Aórtica Transcateter
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
País/Região como assunto:
America do norte
Idioma:
En
Revista:
JACC Cardiovasc Interv
Ano de publicação:
2024
Tipo de documento:
Article