Discrepancies in Sagittal Alignment of the Lower Extremity Among Different Brands of Robotic Total Knee Arthroplasty Systems.
An, Hao-Ming; Wen, Jia-Xin; Gu, Wang; Chen, Ji-Ying; Chai, Wei; Li, Rui.
Afiliação
  • An HM; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China; School of Medicine, Nankai University, Tianjin, China; Chinese PLA Medical School, Beijing, China.
  • Wen JX; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China; School of Medicine, Nankai University, Tianjin, China; Chinese PLA Medical School, Beijing, China.
  • Gu W; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China.
  • Chen JY; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China.
  • Chai W; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China.
  • Li R; Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, Beijing, China; National Clinical Research Center for Orthopedics, Sports Medicine and Rehabilitation, Beijing, China.
J Arthroplasty ; 39(9): 2248-2253, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38508345
ABSTRACT

BACKGROUND:

There is an increasing number of different brands of robotic total knee arthroplasty (TKA) systems. Most robotic TKA systems share the same coronal alignment, while the definitions of sagittal alignment vary. The purpose of this study was to investigate whether these discrepancies impact the sagittal alignment of the lower extremity.

METHODS:

A total of 72 lower extremity computed tomography scans were included in our study, and 3-dimensional models were obtained using software. A total of 7 brands of robotic TKA systems were included in the study. The lower extremity axes were defined based on the surgical guide for each implant. We also set the intramedullary axis as a reference to evaluate the discrepancies in sagittal alignment of each brand of robotic system.

RESULTS:

On the femoral side, the axis definition was the same for all 7 robotic TKA systems. The robotic TKA axes showed a 2.41° (1.58°, 3.38°) deviation from the intramedullary axis. On the tibial side, the 7 robots had different axis definitions. The tibial mechanical axis of 6 of the TKA systems was more flexed than that of the intramedullary axis, which means the posterior tibial slope was decreased while the tibial mechanical axis of the remaining system was more extended.

CONCLUSIONS:

The sagittal alignment of the lower extremity for 7 different brands of robotic TKA systems differed from each other and all deviated from the intramedullary axis. Surgeons should be aware of this discrepancy when using different brands of robotic TKA systems to avoid unexpected sagittal alignment and corresponding adverse clinical outcomes. LEVEL OF EVIDENCE Level IV, Therapeutic Study.