Normative electromyography data and influencing factors in intraoperative neuromonitoring using adhesive skin electrodes during thyroid surgery.
Shin, Sung-Chan; Cheon, Yong-Il; Lee, Minhyung; Sung, Eui-Suk; Lee, Jin-Choon; Kim, Mijin; Kim, Bo Hyun; Lee, Byung-Joo.
Afiliação
  • Shin SC; Department of Otorhinolaryngology-Head and Neck Surgery, College of Medicine, Pusan National University, Busan, Republic of Korea.
  • Cheon YI; Biomedical Research Institute, Pusan National University Hospital, Busan, Republic of Korea.
  • Lee M; Department of Otorhinolaryngology-Head and Neck Surgery, College of Medicine, Pusan National University, Busan, Republic of Korea.
  • Sung ES; Biomedical Research Institute, Pusan National University Hospital, Busan, Republic of Korea.
  • Lee JC; Department of Otorhinolaryngology-Head and Neck Surgery, College of Medicine, Pusan National University, Busan, Republic of Korea.
  • Kim M; Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, Gyeongsangnam-do, Republic of Korea.
  • Kim BH; Department of Otorhinolaryngology-Head and Neck Surgery, College of Medicine, Pusan National University, Busan, Republic of Korea.
  • Lee BJ; Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, Gyeongsangnam-do, Republic of Korea.
Gland Surg ; 13(3): 351-357, 2024 Mar 27.
Article em En | MEDLINE | ID: mdl-38601295
ABSTRACT

Background:

Skin electrodes have been reported to be a useful alternative recording method for intraoperative neuromonitoring (IONM) and show typical electromyography (EMG) waveforms while overcoming the shortcomings of the EMG endotracheal tube. However, the skin electrodes showed relatively lower evoked amplitudes than other recording methods. In this study, we analyzed normative EMG data using skin electrodes and factors that affect the evoked amplitude of thyroid IONM.

Methods:

In total, 167 patients [242 nerves at risk (NAR)] who underwent thyroidectomy under IONM with adhesive skin electrodes were analyzed. A pair of skin electrodes was attached to the lateral border of the lamina of the thyroid cartilage. Evoked EMG data, including mean amplitude and latency, obtained after stimulation of the recurrent laryngeal nerve (RLN) and vagus nerve (VN), were collected and analyzed.

Results:

The mean amplitudes of RLN and VN recorded via skin electrodes were 255.48±96.53 and 236.15±69.72 µV, respectively. The mean latency of the right and left RLN was 3.22±0.03 and 3.49±0.08 mS, respectively. The mean latency of the right and left VN was 5.37±0.80 and 7.57±0.10 mS, respectively. The mean amplitude was significantly lower in the obesity, male, and total thyroidectomy (TT) groups. As body mass index (BMI) and age increased, the amplitude of EMG tended to decrease significantly.

Conclusions:

The evoked amplitude recorded with the skin electrodes was relatively low. A larger surgical extent, obesity, male sex, and age >55 years showed significantly lower evoked amplitudes.