Your browser doesn't support javascript.
loading
Real-Time Gated Proton Therapy: Commissioning and Clinical Workflow for the Hitachi System.
Chen, Hao; Gogineni, Emile; Cao, Yilin; Wong, John; Deville, Curtiland; Li, Heng.
Afiliação
  • Chen H; The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
  • Gogineni E; The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
  • Cao Y; The Ohio State University Medical School, Columbus, Ohio, USA.
  • Wong J; The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
  • Deville C; Dana-Farber/Brigham and Women's Cancer Center, Boston, Massachusetts, USA.
  • Li H; The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
Int J Part Ther ; 11: 100001, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38757076
ABSTRACT

Purpose:

To describe the commissioning of real-time gated proton therapy (RGPT) and the establishment of an appropriate clinical workflow for the treatment of patients. Materials and

Methods:

Hitachi PROBEAT provides pencil beam scanning proton therapy with an advanced onboard imaging system including real-time fluoroscopy. RGPT utilizes a matching score to provide instantaneous system performance feedback and quality control for patient safety. The CIRS Dynamic System combined with a Thorax Phantom or plastic water was utilized to mimic target motion. The OCTAVIUS was utilized to measure end-to-end dosimetric accuracy for a moving target across a range of simulated situations. Using this dosimetric data, the gating threshold was carefully evaluated and selected based on the intended treatment sites and planning techniques. An image-guidance workflow was developed and applied to patient treatment.

Results:

Dosimetric data demonstrated that proton plan delivery uncertainty could be within 2 mm for a moving target. The dose delivery to a moving target could pass 3%/3 mm gamma analysis following the commissioning process and application of the clinical workflow detailed in this manuscript. A clinical workflow was established and successfully applied to patient treatment utilizing RGPT. Prostate cancer patients with implanted platinum fiducial markers were treated with RGPT. Their target motion and gating signal data were available for intrafraction motion analysis.

Conclusion:

Real-time gated proton therapy with the Hitachi System has been fully investigated and commissioned for clinical application. RGPT can provide advanced and reliable real-time image guidance to enhance patient safety and inform important treatment planning parameters, such as planning target volume margins and uncertainty parameters for robust plan optimization. RGPT improved the treatment of patients with prostate cancer in situations where intrafraction motion is more than defined tolerance.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Int J Part Ther Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Int J Part Ther Ano de publicação: 2024 Tipo de documento: Article