The Impact of Back Optic Zone Design in Orthokeratology on Visual Performance.
Transl Vis Sci Technol
; 13(5): 12, 2024 May 01.
Article
em En
| MEDLINE
| ID: mdl-38758570
ABSTRACT
Purpose:
To evaluate the visual performance in adolescents undergoing orthokeratology (OrthoK) treatment with two different optical zone diameters (OZDs).Methods:
This randomized, double-masked, self-controlled prospective study was conducted at Tianjin Eye Hospital (Tianjin, China) in June 2022. Thirty-six eligible schoolchildren were enrolled and fitted with corneal refractive therapy lenses with two sizes of OZDs (5 mm [5OZ] and 6 mm [6OZ]). Each participant was randomized to wear the 5OZ in one eye and the 6OZ in the contralateral eye. Subjective visual quality was assessed using visual acuity, refraction, contrast sensitivity function, and visual symptoms, and the objective optical quality was assessed using ocular higher order aberrations (HOAs) and modulation transfer function (MTF).Results:
Thirty-five myopic children completed a 1-month follow-up visit. The 5OZ lens induced significantly smaller treatment zone diameters than the 6OZ lens (P < 0.001). Subjective visual quality did not differ significantly between the two groups. Compared to baseline, aberrations of Z40, coma-like, spherical-like, and total HOAs in both groups increased significantly (P < 0.05). For the 3-mm pupils, spherical aberration in the 5OZ group was significantly higher than that in the 6OZ group (P < 0.05). The MTF value of the 6OZ group was significantly higher than that of 5OZ group for 0.3 and 1.5 cycles per degree for the 3-mm pupils (P = 0.006 and P = 0.026, respectively). However, HOAs or MTF did not differ significantly between the two groups for the 5-mm pupils.Conclusions:
The difference induced by varying OZD was significant only in the smaller pupil condition. The selection of OZD in OrthoK designs in real-world patient management should be done while considering individual pupil size. Translational Relevance This study revealed that the objective visual quality of small OZD lenses was only slightly affected for the small pupil size.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Refração Ocular
/
Acuidade Visual
/
Procedimentos Ortoceratológicos
/
Miopia
Limite:
Adolescent
/
Child
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
Transl Vis Sci Technol
Ano de publicação:
2024
Tipo de documento:
Article