Your browser doesn't support javascript.
loading
Commissural alignment and the ACURATE neo2 transcatheter aortic valve: Impact on valve performance.
Travieso, Alejandro; Toggweiler, Stefan; Montarello, Nicholas; Renker, Matthias; Tirado-Conte, Gabriela; Loretz, Lucca; Charitos, Efstratios I; Kim, Won-Keun; De Backer, Ole.
Afiliação
  • Travieso A; The Heart Center, Rigshospitalet, Copenhagen, Denmark.
  • Toggweiler S; Heart Center Lucerne, Lucerner Kantonsspital, Lucerne, Switzerland.
  • Montarello N; The Heart Center, Rigshospitalet, Copenhagen, Denmark.
  • Renker M; Department of Cardiology, Kerckhoff Heart Center, Bad Nauheim, Germany.
  • Tirado-Conte G; Department of Cardiac Surgery, Kerckhoff Heart Center, Bad Nauheim, Germany.
  • Loretz L; DZHK (German Center for Cardiovascular Research), Site Rhein-Main, Bad Nauheim, Germany.
  • Charitos EI; The Heart Center, Rigshospitalet, Copenhagen, Denmark.
  • Kim WK; Heart Center Lucerne, Lucerner Kantonsspital, Lucerne, Switzerland.
  • De Backer O; Department of Cardiology, Kerckhoff Heart Center, Bad Nauheim, Germany.
Catheter Cardiovasc Interv ; 104(1): 115-124, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38764320
ABSTRACT

BACKGROUND:

Transcatheter aortic valve replacement (TAVR) is increasingly being used to treat severe aortic stenosis in younger patients. Accordingly, lifetime management regarding future reintervention and coronary access is a concern.

AIMS:

To assess the impact of commissural alignment on ACURATE neo2 transcatheter aortic valve (TAV) performance.

METHODS:

COMALIGN-neo2 was an observational, retrospective study enrolling consecutive TAVR patients treated with the ACURATE neo2 (October 2021 to October 2022). The degree of commissural (mis)-alignment (CMA) with the native aortic valve commissures was determined and transvalvular gradient, effective orifice area, patient-prosthesis mismatch (PPM), and aortic regurgitation (AR) were assessed.

RESULTS:

Among 825 patients, the mean age was 80.7 years and 42% were female. Commissural alignment was achieved in 60% of cases; mild (26%), moderate (9%), and severe misalignment (5%) were found less often. Severe PPM occurred more frequently in patients with severe CMA (14.7%) compared to aligned valves (p = 0.034). By multivariate analysis, severe CMA (odds ratio [OR] 3.12, 95% confidence interval [CI] [1.09-8.90]; p = 0.033) and lack of postdilatation (OR 3.85, [1.33-11.1]; p = 0.012) were associated with severe PPM. Higher rates of ≥mild AR (51.4%) were found in TAVs implanted with severe CMA compared to aligned (34.3%), mildly (38.1%) or moderately (36.0%) misaligned TAVs (p = 0.030). Multivariate analysis identified severe CMA (OR 2.05, [1.05-4.02]; p = 0.037) to be an independent predictor of ≥mild AR.

CONCLUSIONS:

COMALIGN-neo2 is the largest study to date assessing the impact of commissural alignment on acute TAV performance. Severe CMA with the ACURATE neo2 platform was associated with worse valve hemodynamics and increased risk for mild AR.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Insuficiência da Valva Aórtica / Estenose da Valva Aórtica / Desenho de Prótese / Próteses Valvulares Cardíacas / Recuperação de Função Fisiológica / Substituição da Valva Aórtica Transcateter / Hemodinâmica Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Catheter Cardiovasc Interv Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Insuficiência da Valva Aórtica / Estenose da Valva Aórtica / Desenho de Prótese / Próteses Valvulares Cardíacas / Recuperação de Função Fisiológica / Substituição da Valva Aórtica Transcateter / Hemodinâmica Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Catheter Cardiovasc Interv Ano de publicação: 2024 Tipo de documento: Article