Your browser doesn't support javascript.
loading
Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm.
Zhou, Yi; Xia, Jia-Hui; Zhang, Wei; Li, Yan; Wang, Ji-Guang.
Afiliação
  • Zhou Y; Department of Epidemiology and Biostatistics, School of Public Health.
  • Xia JH; Department of Cardiovascular Medicine, Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Zhang W; Department of Cardiovascular Medicine, Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Li Y; Department of Cardiovascular Medicine, Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Wang JG; Department of Epidemiology and Biostatistics, School of Public Health.
Blood Press Monit ; 29(4): 217-220, 2024 Aug 01.
Article em En | MEDLINE | ID: mdl-38774977
ABSTRACT

OBJECTIVE:

We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm.

METHODS:

An approximately equal number of patients with persistent atrial fibrillation and individuals with sinus rhythm were recruited from outpatients and inpatients of Ruijin Hospital, Shanghai, China. BP was measured three times consecutively with a 30-s interval with the OMRON HEM-7361T automatic electronic BP monitor for atrial fibrillation detection. A hand-held single lead electrocardiogram device was used for simultaneous electrocardiogram recordings.

RESULTS:

The device accurately identified atrial fibrillation in 100 (99.0%) of the 101 patients, with only 1 patient incorrectly classified as non-atrial fibrillation. The device correctly identified 99 (95.2%) of the 104 participants with sinus rhythm as non-atrial fibrillation, with five participants incorrectly classified as atrial fibrillation. The device had a positive predictive value of 95.2%, negative predictive value of 99.0%, and overall accuracy of 97.1%. Among the six misclassified participants, one with atrial fibrillation had a heart rate of 65 beats/min, and four of the five participants with sinus rhythm had cardiac arrhythmias (atrial or ventricular premature beat in one participants, sinus tachycardia in one participant, and both arrhythmias in one participant).

CONCLUSION:

The OMRON HEM-7361T BP monitor is accurate in the differentiation between atrial fibrillation and sinus rhythm. Whether the device is sufficiently accurate in the differentiation between atrial fibrillation and other cardiac arrhythmias remains under investigation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2024 Tipo de documento: Article