Efficacy of texture and color enhancement imaging for the visibility and diagnostic accuracy of non-polypoid colorectal lesions.
Article
em En
| MEDLINE
| ID: mdl-38817687
ABSTRACT
Objective:
A newly launched endoscopy system (EVIS X1, CV-1500; Olympus) is equipped with texture and color enhancement imaging (TXI). We aimed to investigate the efficacy of TXI for the visibility and diagnostic accuracy of non-polypoid colorectal lesions.Methods:
We examined 100 non-polypoid lesions in 42 patients from the same position, angle, and distance of the view in three modes white light imaging (WLI), narrow-band imaging (NBI), and TXI. The primary outcome was to compare polyp visibility in the three modes using subjective polyp visibility score and objective color difference values. The secondary outcome was to compare the diagnostic accuracy without magnification.Results:
Overall, the visibility score of TXI was significantly higher than that of WLI (3.7 ± 1.1 vs. 3.6 ± 1.1; p = 0.008) and lower than that of NBI (3.7 ± 1.1 vs. 3.8 ± 1.1; p = 0.013). Color difference values of TXI were higher than those of WLI (11.5 ± 6.9 vs. 9.1 ± 5.4; p < 0.001) and lower than those of NBI (11.5 ± 6.9 vs. 13.1 ± 7.7; p = 0.002). No significant differences in TXI and NBI (visibility score 3.7 ± 1.0 vs. 3.8 ± 1.1; p = 0.833, color difference values 11.6 ± 7.1 vs. 12.9 ± 8.3; p = 0.099) were observed for neoplastic lesions. Moreover, the diagnostic accuracy of TXI was significantly higher than that of NBI (65.5% vs. 57.6%, p = 0.012) for neoplastic lesions.Conclusions:
TXI demonstrated higher visibility than that of WLI and lower than that of NBI. Further investigations are warranted to validate the performance of the TXI mode comprehensively.