High-precision K isotopic analysis of cerebrospinal fluid and blood serum microsamples via multicollector inductively coupled plasma-mass spectrometry equipped with 1013 Ω faraday cup amplifier resistors.
Anal Chim Acta
; 1315: 342812, 2024 Aug 01.
Article
em En
| MEDLINE
| ID: mdl-38879212
ABSTRACT
BACKGROUND:
Potassium isotopic analysis is increasingly performed in both geological and biological contexts as a result of the introduction of MC-ICP-MS instrumentation either equipped with a collision/reaction cell or having the capability of working at "extra-high" mass resolution in order to deal with spectral interference caused by argon hydride (ArH+) ions. Potassium plays an important role in the central nervous system, and its isotopic analysis could provide an enhanced insight into the corresponding processes, but K isotopic analysis of cerebrospinal fluid is challenging due to the small volume, a few microliter only, typically available. This work aimed at developing a method for determining the K isotopic signature of serum and cerebrospinal fluid at a final K concentration of 25 ng mL-1 using Faraday cup amplifiers equipped with a 1013 Ω resistor.RESULTS:
Potassium isotope ratios obtained for reference materials measured at a final K concentration of 25 ng mL-1 were in excellent agreement with the corresponding reference values and the internal and external precision for the δ41K value was 0.11 (2SE, N = 50) and 0.10 (2SD, N = 6), respectively. The robustness against the presence of matrix elements and the concentration mismatch between sample and standard observed at higher K concentrations is preserved at low K concentration. Finally, K isotopic analysis of serum and cerebrospinal fluid (3-12 µL of sample) of healthy mice of both sexes was performed, revealing a trend towards an isotopically lighter signature for serum and cerebrospinal fluid from female individuals, however being significant for serum only.SIGNIFICANCE:
This work provides a robust method for high-precision K isotopic analysis at a concentration of 25 ng mL-1. By monitoring both K isotopes, 39K and 41K, with Faraday cups connected to amplifiers with 1013 Ω resistors, accurate K isotope ratio results are obtained with a two-fold improvement in internal and external precision compared to those obtained with the set-up with traditional 1011 Ω resistors. The difference in the K isotope ratio in CSF and serum between the sexes, is possibly indicating an influence of the sex or hormones on the fractionation effects accompanying cellular uptake/release.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Potássio
/
Espectrometria de Massas
Limite:
Animals
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
Anal Chim Acta
Ano de publicação:
2024
Tipo de documento:
Article