Your browser doesn't support javascript.
loading
High-precision K isotopic analysis of cerebrospinal fluid and blood serum microsamples via multicollector inductively coupled plasma-mass spectrometry equipped with 1013 Ω faraday cup amplifier resistors.
Hobin, Kasper; Costas Rodríguez, Marta; Van Wonterghem, Elien; Vandenbroucke, Roosmarijn E; Vanhaecke, Frank.
Afiliação
  • Hobin K; Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium.
  • Costas Rodríguez M; Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium; Centro de Investigación Mariña, Universidade de Vigo, Departamento de Química Analítica y Alimentaria, Grupo QA2, 36310, Vigo, Spain.
  • Van Wonterghem E; Barriers in Inflammation Lab, VIB Center for Inflammation Research, 9052, Ghent, Belgium; Department of Biomedical Molecular Biology, Ghent University, 9052, Ghent, Belgium.
  • Vandenbroucke RE; Barriers in Inflammation Lab, VIB Center for Inflammation Research, 9052, Ghent, Belgium; Department of Biomedical Molecular Biology, Ghent University, 9052, Ghent, Belgium.
  • Vanhaecke F; Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium. Electronic address: frank.vanhaecke@ugent.be.
Anal Chim Acta ; 1315: 342812, 2024 Aug 01.
Article em En | MEDLINE | ID: mdl-38879212
ABSTRACT

BACKGROUND:

Potassium isotopic analysis is increasingly performed in both geological and biological contexts as a result of the introduction of MC-ICP-MS instrumentation either equipped with a collision/reaction cell or having the capability of working at "extra-high" mass resolution in order to deal with spectral interference caused by argon hydride (ArH+) ions. Potassium plays an important role in the central nervous system, and its isotopic analysis could provide an enhanced insight into the corresponding processes, but K isotopic analysis of cerebrospinal fluid is challenging due to the small volume, a few microliter only, typically available. This work aimed at developing a method for determining the K isotopic signature of serum and cerebrospinal fluid at a final K concentration of 25 ng mL-1 using Faraday cup amplifiers equipped with a 1013 Ω resistor.

RESULTS:

Potassium isotope ratios obtained for reference materials measured at a final K concentration of 25 ng mL-1 were in excellent agreement with the corresponding reference values and the internal and external precision for the δ41K value was 0.11 ‰ (2SE, N = 50) and 0.10 ‰ (2SD, N = 6), respectively. The robustness against the presence of matrix elements and the concentration mismatch between sample and standard observed at higher K concentrations is preserved at low K concentration. Finally, K isotopic analysis of serum and cerebrospinal fluid (3-12 µL of sample) of healthy mice of both sexes was performed, revealing a trend towards an isotopically lighter signature for serum and cerebrospinal fluid from female individuals, however being significant for serum only.

SIGNIFICANCE:

This work provides a robust method for high-precision K isotopic analysis at a concentration of 25 ng mL-1. By monitoring both K isotopes, 39K and 41K, with Faraday cups connected to amplifiers with 1013 Ω resistors, accurate K isotope ratio results are obtained with a two-fold improvement in internal and external precision compared to those obtained with the set-up with traditional 1011 Ω resistors. The difference in the K isotope ratio in CSF and serum between the sexes, is possibly indicating an influence of the sex or hormones on the fractionation effects accompanying cellular uptake/release.