Your browser doesn't support javascript.
loading
Repeatability of a Virtual Reality Headset Perimeter in Glaucoma and Ocular Hypertensive Patients.
Nascimento E Silva, Rafaella; Kim, Julia A; Li, Yihao; Chen, Chen; Chaudhry, Anisa F; Berneshawi, Andrew R; Zhang, Miao; Villarreal, Adam; Liu, Jocelyn; Shue, Ann; Chang, Dolly S; Chang, Robert T.
Afiliação
  • Nascimento E Silva R; Genentech, Inc., South San Francisco, CA, USA.
  • Kim JA; Genentech, Inc., South San Francisco, CA, USA.
  • Li Y; Genentech, Inc., South San Francisco, CA, USA.
  • Chen C; Genentech, Inc., South San Francisco, CA, USA.
  • Chaudhry AF; Byers Eye Institute, Stanford University School of Medicine, Palo Alto, CA, USA.
  • Berneshawi AR; Byers Eye Institute, Stanford University School of Medicine, Palo Alto, CA, USA.
  • Zhang M; Genentech, Inc., South San Francisco, CA, USA.
  • Villarreal A; Genentech, Inc., South San Francisco, CA, USA.
  • Liu J; Byers Eye Institute, Stanford University School of Medicine, Palo Alto, CA, USA.
  • Shue A; Byers Eye Institute, Stanford University School of Medicine, Palo Alto, CA, USA.
  • Chang DS; Genentech, Inc., South San Francisco, CA, USA.
  • Chang RT; Byers Eye Institute, Stanford University School of Medicine, Palo Alto, CA, USA.
Transl Vis Sci Technol ; 13(6): 14, 2024 Jun 03.
Article em En | MEDLINE | ID: mdl-38899952
ABSTRACT

Purpose:

The VisuALL S is an automated, static threshold, virtual reality-based perimeter for mobile evaluation of the visual field. We examined same-day and 3-month repeatability.

Methods:

Adult participants with a diagnosis of glaucoma or ocular hypertension underwent two VisuALL 24-2 Normal T- Full threshold strategy tests at baseline and one additional exam at 3 months for each eligible eye. Spearman, intraclass correlation coefficients (ICCs), and Bland-Altman plots were used to assess the correlation of individual point sensitivities and mean deviation (MD) among three tests.

Results:

Eighty-eight eyes (44 participants) were included. Average age was 68.1 ± 14.3 years, and 60.7% were male. VisuALL MD was highly correlated between tests (intravisit r = 0.89, intervisit r = 0.82; P < 0.001 for both). Bland-Altman analysis showed an average difference in intravisit MD of -0.67 dB (95% confidence interval [CI], -6.04 to 4.71 dB) and -0.15 dB (95% CI, -8.04 to 7.73 dB) for intervisit exams. Eight-five percent of pointwise intravisit ICCs were above 0.75 (range, 0.63 to 0.93), and 65% of pointwise intervisit ICCs were above 0.75 (range, 0.55 to 0.91).

Conclusions:

VisuALL demonstrated high correlation of MD between tests and good repeatability for individual point sensitivities among three tests in 3 months, except at the points around the blind spot and superiorly. Translational Relevance The preliminary reproducibility results for VisuALL are encouraging. Its portable design makes it a potentially useful tool for patients with glaucoma, enabling more frequent assessments both at home and in clinical settings.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Campos Visuais / Glaucoma / Hipertensão Ocular / Testes de Campo Visual / Realidade Virtual Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Transl Vis Sci Technol Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Campos Visuais / Glaucoma / Hipertensão Ocular / Testes de Campo Visual / Realidade Virtual Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Transl Vis Sci Technol Ano de publicação: 2024 Tipo de documento: Article